Lyophilised colourimetric LAMP for visual readout with dual colour indicators
Kenny Malpartida-Cardenas1, Oliver W Stringer1, Matthew L Cavuto1
1Department of Infectious Disease, Imperial College London, London, UK. k.malpartida-cardenas16@imperial.ac.uk.
The Analyst
|March 26, 2026
Summary
Lyophilised colourimetric LAMP (lyo-cLAMP) offers a stable, instrument-free method for pathogen detection. This ready-to-use format eliminates cold-chain requirements, enhancing point-of-care diagnostics.
Area of Science:
- Molecular Diagnostics
- Isothermal Amplification Technologies
Background:
- Loop-mediated isothermal amplification (LAMP) is a viable alternative to PCR for pathogen detection.
- Colourimetric LAMP (cLAMP) enables instrument-free readout, ideal for low-resource settings.
- Current cLAMP formats require cold-chain storage and trained personnel, limiting point-of-care use.
Purpose of the Study:
- To develop and optimize ready-to-use lyophilised colourimetric LAMP (lyo-cLAMP).
- To create a stable, cold-chain-independent cLAMP assay for point-of-care diagnostics.
- To demonstrate the adaptability and accuracy of lyo-cLAMP for pathogen detection.
Main Methods:
- Screening and combining pH and metal indicators for optimal colourimetric discrimination.
- Optimizing lyophilisation protocols for ready-to-use cLAMP formulations.
- Assessing lyo-cLAMP performance using synthetic SARS-CoV-2 RNA and clinical samples.
Main Results:
- Identified four pH and metal indicator combinations with high discrimination.
- Achieved equivalent limit-of-detection for lyo-cLAMP compared to liquid cLAMP with SARS-CoV-2 RNA.
- Demonstrated high accuracy in screening clinical samples for SARS-CoV-2.
Conclusions:
- Developed lyophilised colourimetric LAMP (lyo-cLAMP) as a stable, reagent-free, cold-chain-independent diagnostic tool.
- lyo-cLAMP simplifies molecular diagnostics for point-of-care applications.
- The assay is adaptable to various targets, crucial for rapid response during outbreaks.
Related Concept Videos
Photoluminescence: Applications
1.2K
Photoluminescence offers a wide range of applications due to its inherent sensitivity and selectivity. This technique allows for both direct and indirect analyses of the analyte. Direct quantitative analysis is possible when the analyte exhibits a favorable quantum yield for fluorescence or phosphorescence. However, an indirect analysis may be feasible if the analyte is not fluorescent or phosphorescent, or if the quantum yield is unfavorable. Indirect methods include reacting the analyte with...
1.2K
UV–Vis Spectrometers
4.5K
The absorbance of UV and visible (UV–visible) radiations is measured using a UV–visible spectrophotometer. Deuterium lamps, which emit UV radiation, and tungsten lamps, which produce radiation in the visible region, are used as light sources in UV–visible spectrophotometers. A monochromator or prism is used for diffraction grating, i.e., to split the incoming radiation into different wavelengths. A system of slits is used to focus the desired wavelength on the sample cell.
4.5K
Fluorescence and Phosphorescence: Instrumentation
1.9K
Fluorometers and spectrofluorometers are two types of instruments used for measuring molecular fluorescence. These instruments differ in how they select excitation and emission wavelengths and the type of light sources they utilize. Fluorometers use absorption interference filters to choose excitation and emission wavelengths. The excitation source in a fluorometer is typically a low-pressure mercury vapor lamp that emits intense lines distributed throughout the ultraviolet and visible regions.
1.9K