A Numerical Alternative to MR Thermometry for Safety Validation of Multi-Channel RF Transmit Coils
Alireza Sadeghi-Tarakameh1, Simon Schmidt1, Matt Waks1
1Center for Magnetic Resonance Research (CMRR), University of Minnesota, Minneapolis, Minnesota, USA.
Purpose:
This study proposes a numerical technique to estimate peak local specific absorption rate (SAR) uncertainty of multi-channel RF coils during the process of safety validation as an alternative to experimental temperature and electric-field measurements, and demonstrates its use to enable human studies at 10.5 T.
Methods:
To ensure patient safety, SAR limits established under international guidelines must not be exceeded. Predicting SAR on state-of-the-art parallel transmit systems relies on electromagnetic simulations, which require extensive experimental validation. Despite a well-established validation workflow, SAR prediction errors are unavoidable and must be quantified as a safety margin. While MRT tests are commonly used for this purpose, their technical challenges necessitate an alternative. The proposed technique propagates the error between experimentally and numerically acquired distributions to the uncertainty in simulated peak local SAR using Monte-Carlo simulations without the need for MRT. This method was validated using a 16-channel transceiver 10.5 T torso coil, as well as an 8-channel 10.5 T head coil.
Results:
The proposed numerical technique proved more conservative than existing MRT-based SAR error quantification methods across all tested scenarios. Its application to validate three state-of-the-art head coils (16Tx/32Rx, 16Tx/80Rx, and 16Tx/128Rx) led to regulatory approval for human head imaging and high-quality diffusion and functional MRI results at 10.5 T.
Conclusion:
The proposed technique requires only the experimental acquisition of maps for comparison with simulations, enabling the estimation of SAR prediction uncertainty. This technique was applied to three 16-channel transmit arrays, each used in conjunction with high-channel-count receive arrays for in vivo imaging.
More Related Videos
10:52Design, Instrumentation and Usage Protocols for Distributed In Situ Thermal Hot Spots Monitoring in Electric Coils using FBG Sensor Multiplexing
Published on: March 8, 2020
06:17Quantifying the Relative Thickness of Conductive Ferromagnetic Materials Using Detector Coil-Based Pulsed Eddy Current Sensors
Published on: January 16, 2020
Related Concept Videos
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Mutual Inductance
When two circuits carrying time-varying currents are close to one another, the magnetic flux through each circuit varies because of the changing current in the other circuit. Consequently, an emf is induced in each circuit by the changing current in the other. Therefore, this type of emf is called...
Generating Electromagnetic Radiations
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Applications of EMF Measurements
Equivalent Resistance