Retrospective evaluation of mandibular third molars using simulated low-dose cone-beam computed tomography: A
Dalia Homar Asan1, Josefine Cederhag1, Xie-Qi Shi2
1Department of Oral and Maxillofacial Radiology, Malmö University, Malmö, Sweden.
Imaging Science in Dentistry
|April 3, 2026
Summary
Researchers developed a method using ImageJ to simulate low-dose cone-beam computed tomography (CBCT) scans. Simulated low-dose CBCT images showed comparable diagnostic accuracy to clinical scans, offering a radiation-free simulation approach.
Area of Science:
- Medical Imaging
- Radiology
- Dental Imaging
Background:
- Cone-beam computed tomography (CBCT) is crucial in dentistry.
- Reducing radiation dose in CBCT is a significant clinical goal.
- Simulating low-dose CBCT images can aid in protocol optimization without patient exposure.
Purpose of the Study:
- To establish a procedure for simulating low-dose CBCT scans using ImageJ.
- To evaluate the diagnostic accuracy of simulated low-dose CBCT images compared to clinical low-dose CBCT scans.
- To assess the feasibility of using sinogram data manipulation for radiation-free image simulation.
Main Methods:
- CBCT scans acquired at 5 mA were manipulated using ImageJ to simulate 2 mA settings by adding noise to sinogram data.
- Reconstructed images were compared to original clinical 2 mA CBCT scans.
- Four observers assessed image quality using a ranking scale, with statistical analysis via Wilcoxon signed-rank test and Cohen kappa for intra-observer agreement.
Main Results:
- No significant differences in image quality were found between simulated and clinical low-dose CBCT scans.
- Acceptable intra-observer agreement was achieved, with high agreement in 5 comparisons.
- The evaluation focused on the diagnostic accuracy for mandibular third molars.
Conclusions:
- ImageJ can effectively generate simulated low-dose CBCT scans.
- Simulated low-dose CBCT images demonstrate comparable diagnostic quality to clinical low-dose scans.
- Sinogram data manipulation presents a promising radiation-free method for simulating low-dose CBCT images, aiding in optimization efforts.