Functional Classification of a New IOL Into the Category of Partial Depth of Field: Enhanced IOLs According to the
Giacomo Savini1, Alice Galzignato2, Catarina P Coutinho2,3
1Istituto di Ricovero e Cura a Carattere Scientifico (IRCCS) Bietti Foundation, Rome, Italy.
Purpose:
To assess whether the Evolux (SIFI) intraocular lens (IOL) can be included in the category of partial-depth of field (DOFi) enhanced IOLs according to the functional classification of the European Society of Cataract and Refractive Surgeons.
Methods:
In this prospective study, a consecutive series of patients implanted with the Evolux in one eye was analyzed and compared to a control group with a monofocal IOL (AcrySof SA60WF; Alcon Laboratories, Inc). Preoperative examinations included optical biometry (IOLMaster 700; Carl Zeiss Meditec AG) and anterior segment imaging (MS-39; CSO). At 2 to 3 months postoperatively, monocular corrected distance visual acuity, distance-corrected intermediate visual acuity (DCIVA), distance-corrected near visual acuity (DCNVA), monocular DOFi in 0.2 logarithm of the minimum angle of resolution (logMAR), and the monocular defocus curve were measured in patients implanted with both IOLs.
Results:
The Evolux group included 32 eyes (32 patients) and the monofocal group included 27 eyes (27 patients). The mean DCIVA was, respectively, 0.19 ± 0.11 and 0.30 ± 0.13 logMAR (P = .0009). The mean DCNVA was 0.39 ± 0.12 logMAR in the Evolux group and 0.53 ± 0.18 logMAR in the monofocal group (P = .0007). The mean DOFi in 0.2 logMAR provided by the Evolux was 1.42 ± 0.56 diopters (D); the corresponding value for the monofocal IOL was 1.04 ± 0.54 D (P = .0091). In eyes with the Evolux, the DOFi in 0.2 logMAR DCIVA was negatively correlated with photopic pupil size (r = -0.3748, P = .0346).
Conclusions:
The Evolux offers better visual acuity at an intermediate distance and can be classified as a partial-DOFi enhanced IOL.
More Related Videos
04:59Author Spotlight: Enhancing Visual Outcomes in Cataract Surgery: A Novel Technique to Prevent Posterior Capsular Opacification Through IOL Rotation
Published on: July 7, 2023
07:12Spatio-Temporal In Vivo Imaging of Ocular Drug Delivery Systems using Fiberoptic Confocal Laser Microendoscopy
Published on: September 27, 2021
Related Concept Videos
Focusing of Light in the Eye
Ophthalmic Drug Delivery Systems