Evaluating the Volta phase plate for improved tomogram alignment in cryo-electron tomography
Joshua Hutchings1, Daniel Ji1, Shawn Zheng1
1BioHub, Redwood City, California, USA.
Iucrj
|April 9, 2026
Summary
The Volta phase plate (VPP) improves contrast in cryo-electron tomography (cryo-ET) and enhances tilt-series alignment, crucial for visualizing cellular structures. This makes VPP valuable for studying complex biological samples.
Area of Science:
- Cryo-electron microscopy and tomography
- Structural biology
- Cellular imaging
Background:
- The Volta phase plate (VPP) enhances low-frequency phase contrast in cryo-electron microscopy (cryo-EM) and cryo-electron tomography (cryo-ET).
- Its effect on tomogram alignment, critical for 3D reconstruction, has not been thoroughly investigated.
Purpose of the Study:
- To systematically assess the impact of the VPP on tomogram alignment in cryo-ET.
- To evaluate the VPP's utility for imaging thick and crowded biological specimens.
Main Methods:
- Utilized sub-2 Å single-particle analysis (SPA) and analytical contrast transfer function (CTF) modeling.
- Acquired and compared cryo-ET datasets with and without the VPP.
- Quantitatively analyzed tilt-axis orientation measurements and local tilt-series alignment.
Main Results:
- Despite high-frequency information loss, the VPP enhances low-frequency signal crucial for alignment.
- VPP tomograms show increased contrast, even after denoising, for challenging specimens.
- Improved tilt-axis orientation measurements and a higher success rate for local tilt-series alignment were observed with the VPP.
Conclusions:
- The VPP significantly improves contrast and tilt-series alignment robustness in cryo-ET.
- These benefits are vital for studying cellular ultrastructure and protein organization in crowded cellular environments.
- While VPP may lower resolution for subtomogram averaging, its alignment advantages are critical for moderate-resolution structural studies.
Related Concept Videos
Electron Microscope Tomography and Single-particle Reconstruction
3.1K
Transmission electron microscopy (TEM) can be used to determine the 3D structure of biological samples with the help of techniques such as electron microscope tomography and single-particle reconstruction. While single-particle reconstruction can examine macromolecules and macromolecular complexes in vitro conditions only, tomography permits the study of cell components or small cells in vivo.
Electron Tomography
Electron tomography can be performed either in TEM or STEM (scanning transmission...
Electron Tomography
Electron tomography can be performed either in TEM or STEM (scanning transmission...
3.1K
Cryo-electron Microscopy
4.6K
Conventional electron microscopy (EM) involves dehydration, fixation, and staining of biological samples, which distorts the native state of biological molecules and results in several artifacts. Also, the high-energy electron beam damages the sample and makes it difficult to obtain high-resolution images. These issues can be addressed using cryo-EM, which uses frozen samples and gentler electron beams. The technique was developed by Jacques Dubochet, Joachim Frank, and Richard Henderson, for...
4.6K