Influence of slice thickness on computed tomography perfusion-derived ischemic core volume accuracy in acute ischemic
Chuluunbaatar Otgonbaatar1,2, Huijin Song3, Keun-Hwa Jung4
1Medical Imaging AI Research Center, Canon Medical Systems Korea, Seoul, Republic of Korea.
Background:
Computed tomography (CT) perfusion enables quantitative assessment of cerebral hemodynamic and physiologic parameters and is widely employed for triaging patients with acute ischemic stroke; however, the absence of standardized acquisition protocols for ischemic core estimation remains a major limitation. This study aimed to determine the optimal CT slice thickness for quantifying ischemic core volume derived from CT perfusion in patients with acute ischemic stroke, using diffusion-weighted imaging (DWI)-derived ischemic core volume as the reference standard.
Methods:
In this retrospective study, patients with acute ischemic stroke who met the inclusion criteria were reviewed. CT perfusion data were reconstructed with slice thicknesses of 1, 3, 5, 7, and 10 mm to evaluate their impact on ischemic core volume quantification. Ischemic core volumes were calculated using fully automated post-processing software. The ischemic core was defined using thresholds of relative cerebral blood flow (rCBF) <20-38% and relative cerebral blood volume (rCBV) <34-42%. The optimal slice thickness was identified based on the lowest mean absolute error (MAE), the smallest bias and narrowest 95% limits of agreement (LoA) in Bland-Altman analysis, and the highest intraclass correlation coefficient (ICC) compared with DWI-derived ischemic core volume.
Results:
A total of 87 patients (mean age ± standard deviation, 69.64±12.27 years; 49 males) were included. Among the different slice thicknesses, the lowest MAE was observed at the rCBF <38% threshold compared to rCBF <30%. Bland-Altman analysis and ICC results indicated that 3 mm (bias, 5.04 mL; 95% LoA, -35.29 to 45.38; ICC =0.961), 5 mm (bias, 4.55 mL; 95% LoA, -37.41 to 46.51; ICC =0.961), and 7 mm (bias, 4.75 mL; 95% LoA, -35.14 to 44.66; ICC =0.961) slice thicknesses yielded the most accurate estimates of ischemic core volume when using rCBF <38%. In contrast, 1 mm (bias, 6.71 mL; 95% LoA, -36.86 to 50.29; ICC =0.951) and 10 mm (bias, 9.12 mL; 95% LoA, -47.42 to 65.67; ICC =0.905) reconstructions demonstrated lower agreement with DWI-derived measurements.
Conclusions:
CT perfusion data reconstructed at slice thicknesses of 3-7 mm provided the most accurate estimation of ischemic core volume, at both commonly used and optimal perfusion thresholds, compared with thinner (1 mm) or thicker (10 mm) slice thicknesses.
More Related Videos
06:45Author Spotlight: Integrated Photoacoustic, Ultrasound, and Angiographic Tomography (PAUSAT) for NonInvasive Whole-Brain Imaging of Ischemic Stroke
Published on: June 2, 2023
13:16Intravascular Perfusion of Carbon Black Ink Allows Reliable Visualization of Cerebral Vessels
Published on: January 4, 2013
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies VII: Vascular Imaging
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Cardiac Output II: Effect of Stroke Volume on Cardiac Output
Preload
Preload refers to the initial elongation of the cardiac myocytes before contraction and is related to the volume of blood filling the heart at the end of diastole, or end-diastolic volume. The...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...