Related Experiment Video
Updated: Apr 17, 2026

Optimization of Transesophageal Atrial Pacing to Assess Atrial Fibrillation Susceptibility in Mice
Published on: June 29, 2022
[Development of an Automated Test System for Pacing Performance]
Ya Chen1, Yumin Fan2, Xiaoqing Zhang2
1Shanghai Institute of Medical Device Testing, Shanghai, 201318.
Abstract:
This study presents the design of an automated control system based on LabVIEW development technology, aiming to develop a high-precision and fully automated pacing performance testing system. The system enables automatic control of parameter configuration, test item selection and so on. It significantly improves testing efficiency while ensuring measurement accuracy, and it provides an efficient testing platform for type inspections, thereby promoting the rapid and orderly development of the domestic cardiac pacemaker industry and accelerating the localization of high-end medical devices.