Quantitative Assessment of the Effects of Scatter Radiation on Image Quality During Extremity Radiography at a
Erica Mawusinu Domi1, Jacob Leonard Ago1,2, William K Antwi1
1Department of Radiography, School of Biomedical and Allied Health Sciences University of Ghana Accra Ghana.
Background And Aims:
There is a potential increase in scatter radiation during extremity radiography, where radiographers may divide the detector to obtain multiple images on one film. This study assesses the impact of scatter radiation on image quality (IQ) during the dividing receptor technique in extremity radiography at the Korle-Bu Teaching Hospital, Ghana.
Methods:
An experimental study design was employed. The examinations were performed with a computed radiography system from June 6, 2024, to July 4, 2024, using a dividing receptor technique. Scatter radiation was evaluated as whole-body equivalent dose and surface dose at depths of 10 mm [Hp(10)] and 0.07 mm [Hp(0.07)], respectively. IQ was assessed using ImageJ software v1.54c. Data were analyzed descriptively and inferentially with IBM SPSS v29.0. A p-value < 0.05 was considered statistically significant.
Results:
The measured scatter doses at Hp(10) and Hp(0.07) for all the examinations were 0.017 ± 0.0015 mSv and 0.0016 ± 0.0020 mSv, respectively. Whereas no statistically significant differences in signal-to-noise ratio (p = 0.53) were found for examinations performed before and after dividing the receptor, the difference in noise was found to be statistically significant (p < 0.001). Different anatomical regions generated varying, but non-significant (p > 0.05), scatter radiations, noise, and signal-to-noise ratios. Pearson correlation tests revealed no statistically significant relationship between the reduced IQ and the measured Hp(10) (p = 0.98) and Hp(0.07) (p = 0.85).
Conclusion:
The study identified significant differences in noise and non-significant differences in SNR, although the differences did not result from the scatter dose produced from the dividing receptor technique during extremity radiography. Nonetheless, there is a need for radiographers to adopt strategies to reduce the scatter radiation when using the dividing receptor technique. Consequently, we recommend radiation dose optimization techniques such as effective collimation and protocol adjustments during extremity radiography.
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
Related Concept Videos
Biological Effects of Radiation
X-ray Imaging
Radiological Investigation I: X-ray and CT
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...