Correcting 129Xe Gas Exchange MRI for Incidental Gas-Phase Excitation-Comparing Approaches, and Identifying
Suphachart Leewiwatwong1, Matthew Willmering2,3, Andrew D Hahn4
1Biomedical Engineering, Duke University, Durham, North Carolina, USA.
Purpose:
Incidental excitation of the gas-phase resonance during 129Xe gas exchange MRI introduces contamination that can bias membrane uptake and red blood cell (RBC) transfer metrics. Here, we defined the level of gas-phase contamination that is acceptable, and compared the dual-echo (Hahn) and spectroscopy-informed (Willmering) frameworks to correct for it.
Methods:
Nine healthy volunteers underwent interleaved 3D radial dual-echo Dixon-based 129Xe MRI and spectroscopy. The dissolved-phase RF pulse was shortened from its optimal value to increase incidental gas contamination. The Hahn and Willmering corrections were implemented and refined using a grid search over a contamination-dominated ROI to determine any platform-specific scale and phase adjustments. The effects of contamination phase and scale were tested by synthetically reintroducing gas signal at phase offsets that maximized defect- and high intensity-like artifacts. Membrane/gas and RBC/gas defect and high percentages were evaluated against thresholds derived from a healthy reference cohort to determine acceptable gas contamination.
Results:
Images acquired with the shortened pulse contained 24% ± 6% gas contamination. After applying empirically calibrated phase shifts (Hahn≈230°, Willmering≈49°) and no scale adjustment (optimal value≈1), both correction methods increased RBC SNR by ∼9% and reduced background noise. Phase- and scale-sweep experiments indicated that gas contamination becomes unacceptable once it exceeds ∼9% ± 3%, at which point membrane and RBC defect or high percentages surpass their thresholds.
Conclusions:
Gas-phase contamination beyond ∼9% requires correction. After a one-time empirical phase calibration, this can be equally well achieved by both the Hahn and Willmering methods.
Related Concept Videos
NMR Spectrometers: Resolution and Error Correction
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Gas Exchange and Transport
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...