Influence of body size on the accuracy of lumbar bone mineral density by dual-energy CT: A phantom study
Narumi Nonoura1, Minoru Yamada2, Naoki Nagasawa3
1Department of Radiological Technology, Saitama City Hospital, 2460 Mimuro, Midori-ku, Saitama-shi, Saitama 336-8522, Japan; Division of Health Science, Graduate School of Health Science, Suzuka University of Medical Science, 1001-1 Kishioka, Suzuka-shi, Mie 510-0293, Japan.
Purpose:
To elucidate the relationship between lumbar bone mineral density (BMD) measurements obtained using dual-energy X-ray absorptiometry (DXA) and dual-energy computed tomography (DECT) and to compare the impact of body size on the precision of BMD measurements obtained using both DXA and DECT.
Methods:
Abdominal phantoms simulating four body sizes and containing four bone-equivalent phantoms with varying CaCO3 densities were used. The phantoms were scanned using DXA and DECT at fixed imaging conditions. The relationship between the DXA-measured BMD and the DECT CT value difference (CT value of low tube voltage - CT value of high tube voltage) for the small (S) size phantom was estimated using simple regression analysis. The mean DXA-derived BMD of the S-size phantom was used as the reference value. The effect of four body sizes on the precision of DECT and DXA BMD was measured using the regression equation and compared with the acceptable range of 5.3% least significant change (LSC).
Results:
The coefficient of determination of simple regression analysis between DXA-measured BMD and DECT CT value difference demonstrated a remarkably strong linear relationship (R2 = 0.993). DXA accurately measured BMD up to body size medium within 5.3% LSC but tended to underestimate it for body sizes large (L) and double large (LL). On the other hand, DECT accurately measured BMD up to body size L within 5.3% LSC but tended to overestimate BMD at body size LL.
Conclusion:
DXA and DECT BMD measurements exhibited a strong linear correlation. Body size had a lower influence on the accuracy of BMD measurement obtained using DECT than using DXA, demonstrating higher precision even for larger body sizes.
More Related Videos
Related Concept Videos
Bone Disorders
Bone deposition is also affected by the levels of sex hormones like estrogen and testosterone that promote osteoblast activity and bone matrix synthesis. When the level of these hormones decreases due to aging, it causes a reduction in bone deposition. As a result, bone resorption by osteoclasts...
Drug Dosing: Obese Patients
Changes in the Appendicular Skeleton with Age
Initially, the limb buds consist of a core of mesenchyme covered by a layer of ectoderm. The ectoderm at the end of the limb bud thickens to form a narrow crest called the apical ectodermal ridge. This ridge stimulates the underlying...
Study Designs in Epidemiology
Observational studies are those where the researcher does not intervene but rather observes natural variations. They include cross-sectional, cohort, and...