Comparative evaluation of HbA1c in point-of-care testing (POCT) devices: enzymatic and immunoassay methods compared
Windradini Rahvian Aridama1, Ferdy Royland Marpaung2,3, Yessy Puspitasari2,3
1Study Program of Clinical Pathology, Universitas Airlangga, Surabaya, Indonesia.
Aim:
To evaluate 3 point-of-care testing (POCT) devices in testing HbA1c: the Simplex TASTM101 (Tascom Co. Ltd., South Korea), which utilizes an enzymatic method, and two immunoassay-based devices, the Zybio Q8 Pro and Zybio EXR 110 (Zybio Inc., China).
Methods:
A total of 40 whole blood samples with EDTA anticoagulant were collected patients undergoing HbA1c testing at the outpatient department. The samples were analyzed using 3 POCT devices and one reference method.
Results:
Passing-Bablok regression revealed no significant fixed or proportional bias. Strong correlations with the gold standard were observed for Simplex TASTM101 (r=0.944), Zybio Q8 Pro (r=0.910), and Zybio EXR 110 (r=0.839), all with p<0.001. Bland-Altman analysis confirmed that the majority of measurements fell within an acceptable clinical range.
Conclusion:
This study demonstrated that the three POCT devices employing enzymatic and immunoassay methods served as reliable alternatives to ion exchange-high-performance liquid chromatography (IE-HPLC) for HbA1c measurement in clinical practice.
More Related Videos
07:24A Precision Medicine Tool for Measurement and Monitoring of Hemoglobin S in Sickle Cell Disease Patients Receiving Transfusion Therapy
06:50A High-Throughput Electrochemiluminescence 7-Plex Assay Simultaneously Screening for Type 1 Diabetes and Multiple Autoimmune Diseases
Published on: May 29, 2020
Related Concept Videos
Enzyme-Linked Immunosorbent Assay
There are many different types of ELISAs, but they all involve an antibody molecule whose constant region binds an enzyme, leaving the variable region free to bind its specific antigen. Enzyme-substrate reaction allows the antigen to be visualized or...
High-Performance Liquid Chromatography: Types of Detectors