Related Experiment Video
Updated: May 1, 2026

The Use of Mixed Reality in Custom-Made Revision Hip Arthroplasty: A First Case Report
Published on: August 4, 2022
Surgical repositioning of a SING IMT following haptic dislocation: Restoration of alignment and visual function
Shervine Ameli1,2, Sorcha Ní Dhubhghaill1,2, Faustino Vidal-Aroca3
1Universitair Ziekenhuis Brussel, Laarbeeklaan 101, Jette, 1090, Belgium.
Purpose:
The Smaller-Incision New-Generation Implantable Miniature Telescope (SING IMT) is an IOL designed to enhance central vision in patients with bilateral late-stage AMD. This report describes a case of successful surgical repositioning of a tilted SING IMT following haptic displacement out of the capsular bag.
Observations:
A 77-year-old man with bilateral geographic atrophy and baseline CDVA (Corrected Distance Visual Acuity) of logMAR 0.96 (Snellen 20/200) underwent uncomplicated implantation of a SING IMT in the right eye. Six weeks postoperatively, CDVA showed no improvement, and the patient reported monocular diplopia with off-axis visual perception. Retinal morphology remained stable. Slit-lamp and anterior segment OCT revealed pronounced optic tilt caused by inferior haptic dislocation out of the capsular bag. The haptic was successfully repositioned by performing superior traction with a push-pull hook, resulting in restoration of straight, diplopia-free vision and an improvement in CDVA to logMAR 0.52 (Snellen 20/63).
Conclusions And Importance:
Prompt surgical correction of haptic displacement may fully restore the optical and functional benefits of the SING IMT without compromising corneal health.