Related Experiment Videos

The pneumatonograph. A laboratory study.

R A Moses, W J Grodzki

    Archives of Ophthalmology (Chicago, Ill. : 1960)
    |March 1, 1979
    PubMed
    Summary
    This summary is machine-generated.

    The pneumatonograph (PTG) overestimates intraocular pressure due to scale compression, especially at lower pressures. Its position-sensitivity and scale compression make it unsuitable for tonography.

    Related Concept Videos

    You might also read

    Related Articles

    Articles linked to this work by shared authors, journal, and citation graph.

    Sort by
    Same author

    Male-biased reproduction and sex-ratio adjustment in muskrats.

    Oecologia·2017
    Same author

    Host preferences and temporal trends of the tick Ixodes angustus in north-central Alberta.

    The Journal of parasitology·1998
    Same author

    Pilocarpine-induced movement of the trabecular mesh.

    Australian and New Zealand journal of ophthalmology·1986
    Same author

    The effect of iloprost on the rabbit eye.

    Journal of ocular pharmacology·1986
    Same author

    Pseudofacility. Where did it go?

    Archives of ophthalmology (Chicago, Ill. : 1960)·1985
    Same author

    Mechanism of glaucoma secondary to increased venous pressure.

    Archives of ophthalmology (Chicago, Ill. : 1960)·1985
    Same journal

    The economics of the initial preventive physical examination in medicare-reply.

    Archives of ophthalmology (Chicago, Ill. : 1960)·2013
    Same journal

    Modification of silicone oil retention sutures in aphakic eyes with iris loss-reply;.

    Archives of ophthalmology (Chicago, Ill. : 1960)·2013
    Same journal

    December 2011 archives web quiz winner.

    Archives of ophthalmology (Chicago, Ill. : 1960)·2013
    Same journal

    Angle involvement and glaucoma in patients with biopsy-proven iris melanoma: a response-reply.

    Archives of ophthalmology (Chicago, Ill. : 1960)·2013
    Same journal

    About this journal.

    Archives of ophthalmology (Chicago, Ill. : 1960)·2013
    Same journal

    In memoriam: goodwin m. Breinin, MD (1918-2011).

    Archives of ophthalmology (Chicago, Ill. : 1960)·2013
    See all related articles

    Area of Science:

    • Ophthalmology
    • Biomedical Engineering
    • Medical Devices

    Background:

    • The pneumatonograph (PTG) is a device used for measuring intraocular pressure (IOP).
    • Accurate IOP measurement is crucial for diagnosing and managing conditions like glaucoma.
    • Previous studies have not fully addressed the PTG's limitations regarding scale compression and position-sensitivity.

    Purpose of the Study:

    • To investigate the accuracy and reliability of the pneumatonograph (PTG) for measuring intraocular pressure (IOP).
    • To determine the impact of application force, device orientation, and added weight on PTG readings.
    • To assess the suitability of the PTG for tonography.

    Main Methods:

    • Analysis of PTG scale readings under varying application forces.

    Related Experiment Videos

  • Evaluation of PTG performance when oriented vertically.
  • Assessment of PTG readings with an added 10-g plunger load, as used in tonography.
  • Main Results:

    • PTG scale readings are dependent on the force applied to the cornea.
    • Scale compression occurs at lower IOPs, leading to overestimated pressure readings.
    • Vertical orientation and added plunger weight accentuate scale compression, further compromising accuracy.

    Conclusions:

    • The PTG scale is subject to compression, rendering it inaccurate.
    • The PTG's readings are sensitive to its orientation, impacting IOP measurements.