Richard D Branson1, Thomas C Blakeman, Laura Moore
1Department of Surgery (R.D.B., T.C.B., D.G., D.R., M.G.), Division of Trauma & Critical Care, University of Cincinnati, Cincinnati, Ohio; Department of Surgery (L.M.), University of Texas at Houston, Houston, Texas; Department of Surgery (M.L.), University of Southern California, LA County, Los Angeles, California; Department of Surgery (B.B.), Regions Hospital, Minneapolis, Minnesota; Department of Surgery (J.A.J.), Uniformed Services University, Bethesda, Maryland; Center for Biostatistics (C.M.), Cincinnati Children's Hospital, Cincinnati, Ohio; Research Health Science Officer CSTARS (D.R.), Cincinnati, Ohio.
Background:
Automated physiologic closed-loop control (PCLC) of inspired oxygen fraction (FIO2) may improve oxygen titration while reducing clinician workload. We conducted a randomized noninferiority trial comparing PCLC with protocolized manual control (MC) in mechanically ventilated trauma and acute care surgery patients.
Methods:
In this randomized, controlled, multicenter trial, adult mechanically ventilated patients were randomized to PCLC or MC FIO2 titration targeting arterial oxygen saturation (SpO2) of 94%±2%. The primary effectiveness outcome was the percentage of time SpO2 remained within the target range. The primary safety outcome was hypoxemia (SpO2 <88% at any time). A noninferiority margin of 10% was prespecified. Secondary outcomes included oxygen utilization and clinician interventions. Analyses were performed using a modified intention-to-treat approach.
Results:
One hundred ninety-five subjects completed the trial (98 MC and 97 PCLC). The median time within the SpO2 target range was 36.8% (IQR, 22.1-49.6) with MC and 30.1% (IQR, 18.4-44.7) with PCLC, meeting criteria for noninferiority [the mean±SD of the relative duration on the transformed scale was 0.53 (0.37) and 0.61 (0.36). The mean difference between the transformed data was -0.08 with a one-sided lower 97.5% CL of -0.18. Comparing this lower 95% CL with the prespecified noninferiority margin (NM) of -0.35, this difference lay within the NM. Hypoxemia occurred in fewer patients assigned to PCLC compared with MC (22% vs. 38%, p=0.04). PCLC was associated with lower FIO2 exposure, more frequent automated adjustments, and fewer manual FIO2 adjustments.
Conclusions:
In mechanically ventilated trauma and acute care surgery patients, PCLC of FIO2 was noninferior to protocolized MC for maintaining target oxygen saturation. PCLC was associated with fewer hypoxemic events, reduced oxygen usage, and more frequent automated FIO2 changes. (J Trauma Acute Care Surg. 2026;000:000-000. Copyright © 2026 The Author(s). Published by Wolters Kluwer Health, Inc. on behalf of the American Association for the Surgery of Trauma.).
Level Of Evidence:
Randomized Controlled Trial with single blinding; Level II.
More Related Videos
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
05:39Author Spotlight: A Non-Intubated Video-Assisted Thoracoscopic Surgery with Multimodal Analgesia and Sevoflurane Inhalation Anesthesia
Published on: May 26, 2023
Related Concept Videos
Mechanical Ventilation I: Indication and Settings
Physiological Control of Respiration
Breathing, a seemingly passive process, is regulated by the respiratory center in the brainstem. This center coordinates the involuntary control of respirations, which means it occurs without conscious effort, ensuring a smooth and uninterrupted pattern.
Regulation of Ventilation
The body maintains ventilation by monitoring levels of carbon dioxide (CO2), oxygen (O2), and hydrogen ion concentration (pH) in the arterial blood. Among these factors, the level of CO2 plays a crucial...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Cardiopulmonary Resuscitation II: ACLS Airway Management
Acute Respiratory Failure-V
Ensure that patients are monitored continuously for their response to therapy, including changes in...