Visualizing Periampullary Tumors with Intraductal Ultrasound utilizing an Intracardiac Echocardiography Catheter: A
Mark Ramaekers1, Catarina Dinis Fernandes2, Xueting Li2
1Department of Surgery, Catharina Cancer Institute, Catharina Hospital Eindhoven, Eindhoven, Noord-Brabant, The Netherlands; Department of Electrical Engineering, Eindhoven University of Technology, Eindhoven, Noord-Brabant, The Netherlands.
Objective:
Early detection of pancreatic cancer is crucial for survival, but detecting smalllesions remains challenging. Intraductal Ultrasound (IDUS) using intracardiac echocardiography (ICE) catheters for B-mode and Shear-Wave Elastography (SWE) potentially offers improved visualization and characterization of small tumors. This study assesses the feasibility of IDUS using ICE catheters to detect and visualize periampullary tumors in surgically resected specimens.
Methods:
In this two-phase ex-vivo feasibility study, 25 pancreatic specimens were included, of which the first 10 were used to establish and standardize the imaging protocol, followed by technical feasibility evaluation in the remaining 15 specimens. Catheters were introduced into the pancreatic duct, common bile duct, or positioned extraductally to enable tumor visualization with B-mode imaging and shear-wave elastography (SWE). Tumor visualization rates, catheter insertion success, SWE measurements in normal and tumor tissue, and image quality were assessed.
Results:
ICE catheter insertion was successful in 12 of 15 specimens; unsuccessful access was primarily related to large tumor size (>4 cm) or unidentifiable ductal anatomy following surgical resection. However, extraluminal imaging successfully visualized tumors in one of these cases. Median shear-wave speed and elastic modulus for normal pancreatic parenchyma were 1.58 m/s and 7.6 kPa, respectively. SWE measurements in tumor tissue were suboptimal, likely due to ex-vivo tissue variability and catheter strain during repeated use.
Conclusion:
IDUS with ICE is feasible for qualitative B-mode visualization of periampullary tumors and enables SWE assessment of pancreatic parenchyma in an ex-vivo setting. Reliable elastography of tumor tissue remained challenging, indicating the need for further technical refinement and in-vivo validation.
Related Concept Videos
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for...
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion,...