Background:
- Screw loosening is a frequent mechanical issue in implant restorations.
- The interplay between sealing agents and superstructure materials on screw stability needs more research.
Purpose of the Study:
- To investigate how access channel sealing materials and superstructure materials affect abutment-implant screw stability.
- To analyze screw stability after subjecting restorations to thermomechanical cyclic loading.
Main Methods:
- Forty-eight implant-abutment assemblies with monolithic zirconia or Cerasmart crowns were tested.
- Two sealing protocols were used: Polytetrafluoroethylene (PTFE) + composite or polyvinyl siloxane (PVS) putty.
- Assemblies underwent 750,000 cycles of off-axis loading, and reverse torque values (RTV) were measured.
Main Results:
- A significant interaction was found between restorative material and sealing protocol (p=0.0170).
- Sealing protocol significantly affected RTV (p=0.0499), but superstructure material did not (p=0.8368).
- Zirconia crowns with PVS putty showed higher RTV (36.33 ± 4.53 Ncm) than those with PTFE (29.32 ± 6.30 Ncm).
Conclusions:
- Access channel sealing materials influence abutment screw stability in a material-dependent manner.
- Silicone-based sealing (PVS) enhanced torque maintenance in zirconia restorations.
- Resilient materials like Cerasmart may be less affected by sealing material choice regarding preload maintenance.