Radiotherapy in Patients with Cardiac Implantable Devices: A Single-Centre Retrospective Observational Analysis of
Ellen Saghie1, Roshni Manoj1, Lloyd Tudor1
1Leeds Teaching Hospital Trust, Leeds LS1 3EX, UK.
This study found that a local Standard of Practice (SOP) for patients with cardiac implantable electronic devices (CIEDs) undergoing radiotherapy is safe. No significant changes in lead sensing or impedance were observed, supporting the SOP as a viable alternative to existing guidelines.
Area of Science:
- Cardiology
- Oncology
- Medical Devices
Background:
- Increasing use of radiotherapy alongside cardiac implantable electronic devices (CIEDs) necessitates clear management guidelines.
- Ambiguity exists between international and manufacturer guidelines for managing CIED patients during radiotherapy.
- The British Heart Rhythm Society released new guidance in 2025, highlighting the need for safety data.
Purpose of the Study:
- To evaluate the safety and efficacy of a locally implemented Standard of Practice (SOP) for patients with CIEDs undergoing radiotherapy.
- To assess potential impacts of radiotherapy on CIED function and lead integrity.
- To compare the safety of the local SOP against existing international guidelines.
Main Methods:
- Retrospective, single-centre observational study of 320 patients with CIEDs receiving radiotherapy.
- Patients categorized into non-pacemaker-dependent, pacemaker-dependent, and implantable-cardioverter-defibrillator (ICD) groups.
- Lead sensing and impedance values analyzed pre- and post-radiotherapy using paired Student's T-test.
Main Results:
- No statistically significant changes in lead sensing capabilities across all patient groups (p > 0.05).
- No significant changes in lead impedance for ICD and non-pacemaker-dependent groups.
- A statistically significant but not clinically significant reduction in atrial lead impedance was noted in the pacemaker-dependent cohort (p = 0.039).
Conclusions:
- The locally implemented Standard of Practice (SOP) for radiotherapy in CIED patients is safe.
- The SOP demonstrates comparable safety to existing British Heart Rhythm Society guidelines.
- Findings support the adoption of the local SOP as a safe management strategy.
Related Concept Videos
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Cardiac Catheterization IV: Nursing Management