Cone Beam Computed Tomography (CBCT)-Based Online Adaptive Radiotherapy for Patients With Metal Implants: An
Fiona Li1, Sean Tanny2, Wesley Rivais1
1Radiation Oncology, University of Rochester, Rochester, USA.
Patients with metal implants can now undergo advanced online adaptive radiotherapy (oART) using cone beam computed tomography (CBCT). This study shows oART is feasible with metal artifacts, ensuring accurate radiation delivery and acceptable treatment times.
Area of Science:
- Radiation Oncology
- Medical Physics
- Radiotherapy Technology
Background:
- Metal implants (MIs) often exclude patients from cone beam computed tomography (CBCT)-based online adaptive radiotherapy (oART) due to imaging artifact concerns.
- This exclusion limits access to innovative radiotherapy technologies for patients with MIs.
Purpose of the Study:
- To evaluate the institutional experience and feasibility of using an oART workflow in patients with MIs.
- To assess dose accuracy and on-couch treatment time in patients undergoing oART with MIs.
- To investigate the impact of metal artifacts (MAs) on treatment planning and auto-contouring.
Main Methods:
- Retrospective analysis of eight patients (five with MIs, three without) treated on the Ethos system.
- Implementation of dosimetric avoidance through MIs during treatment planning.
- Evaluation of MA override necessity using 3D gamma analysis and dose metrics (Dmax, Dmin, Dmean).
- Assessment of AI-based auto-contouring performance using Dice Similarity Coefficient (DSC) and Maximum Hausdorff Distance (MHD).
Main Results:
- Dosimetric avoidance achieved clinically acceptable plan quality for MIs.
- No significant dose distribution differences were found with or without MA overrides, suggesting overrides are often unnecessary.
- MI and CBCT artifacts impacted auto-contouring but did not impede oART delivery feasibility.
- Contouring times were comparable between patients with and without MIs.
- Good agreement was observed between auto-contoured and manually edited structures, except for the bladder.
Conclusions:
- The proposed workflow demonstrates the feasibility of CBCT-based oART for patients with MIs.
- Institutional experience supports the use of oART in this patient population.
- Further validation may be needed to meet institutional standards and policies.
More Related Videos
09:49Dual-phase Cone-beam Computed Tomography to See, Reach, and Treat Hepatocellular Carcinoma during Drug-eluting Beads Transarterial Chemo-embolization
Published on: December 2, 2013
08:17Dynamic Lung Tumor Tracking for Stereotactic Ablative Body Radiation Therapy
Published on: June 7, 2015
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...