Evaluation of Optimised Implantable Collamer Lens Sizing using an Optical Low-Coherence Reflectometry Biometry Device
Nikhil Jain1, Masara Laginaf2, Romesh Angunawela2
1Faculty of Life Sciences and Medicine, King's College London, UK.
Accurate Implantable Collamer Lens (ICL) sizing is crucial for preventing vision complications. A revised white-to-white (WTW) subtraction of 0.25mm for the LENSTAR 900 (LS900) may improve ICL sizing accuracy and reduce risks associated with vault height.
Area of Science:
- Ophthalmology
- Refractive Surgery
- Biometry
Background:
- Implantable Collamer Lens (ICL) surgery is a viable treatment for ametropia when laser vision correction is unsuitable.
- Precise lens sizing is critical to avoid suboptimal vault heights, which can lead to complications.
- The Online Calculator and Ordering System (OCOS) uses white-to-white (WTW) measurements for ICL sizing, often with a standard subtraction value.
Purpose of the Study:
- To evaluate if a more specific WTW subtraction value for the LENSTAR 900 (LS900) biometry device can enhance ICL sizing accuracy.
- To assess the impact of different WTW subtraction values on ICL vault height outcomes.
Main Methods:
- A retrospective observational cohort study involving 199 eyes from 102 patients undergoing ICL surgery.
- Pre-operative biometry was performed using the LS900, with ICLs sized via the OCOS calculator using a 0.4mm WTW subtraction.
- Vault height was measured at multiple post-operative time points, and simulations were conducted using a 0.25mm WTW subtraction for the LS900.
Main Results:
- 'Ideal' ICL sizing was achieved in 70% of cases at 3 months, decreasing to 67% at 12 months.
- The majority of incorrectly sized lenses were oversized at 3 months but became undersized by 12 months.
- Simulations indicated that a 0.25mm WTW subtraction for LS900 biometry could shift vault heights away from undersizing at 12 months.
Conclusions:
- Current OCOS calculator sizing with a 0.4mm WTW subtraction may lead to oversizing, potentially due to measurement discrepancies.
- An adjusted WTW subtraction of 0.25mm for LS900 biometry shows potential to reduce the incidence of low vaults, a risk factor for cataract development.
- Further real-world studies are necessary to validate the improved accuracy of the 0.25mm WTW subtraction for ICL sizing.