Reconstruction algorithms and arm positioning effects on abdominal CT image quality and radiation dose: a phantom
Han Song Mun1, Sanghyeok Lim2, Shinhyung Kang3
1Department of Radiology, Seoul St. Mary's Hospital, College of Medicine, The Catholic University of Korea, Seoul, Republic of Korea.
European Radiology Experimental
|May 7, 2026
Summary
Arm positioning significantly impacts abdominal CT radiation dose and image quality. Deep learning reconstruction (DLIR) can maintain diagnostic quality when arms cannot be raised, offering a valuable alternative.
Area of Science:
- Medical Imaging
- Radiology
- Computed Tomography
Background:
- Patient positioning is crucial for optimizing radiation dose and image quality in abdominal CT scans.
- Standard protocols often require arms raised (AU), which may not always be feasible for patients.
Purpose of the Study:
- To investigate the influence of different arm positions and reconstruction algorithms on radiation dose and image quality in abdominal CT.
- To assess the effectiveness of deep learning-based reconstruction (DLIR) in mitigating image quality degradation caused by suboptimal arm positioning.
Main Methods:
- Abdominal CT scans were performed on a phantom using various arm positions (arms up, arms down, arms on belly, with/without air cushions) and noise index settings.
- Images were reconstructed using filtered back projection (FBP), iterative reconstruction, and DLIR.
- Radiation dose and quantitative/qualitative image quality metrics were evaluated.
Main Results:
- Non-arms-up positions significantly increased radiation dose (up to 83%) compared to the arms-up position.
- Deep learning-based reconstruction (DLIR) substantially improved image quality in non-arms-up positions, approaching diagnostic quality.
- Air cushions offered modest, configuration-dependent benefits, more apparent in arms-on-belly positions.
Conclusions:
- Arm positioning is the primary factor affecting radiation dose and image quality in abdominal CT.
- DLIR is effective in preserving diagnostic image quality when patients cannot raise their arms.
- Air cushions provide limited benefits, and arms-on-belly positioning should be avoided if possible.