Evaluation of Commercial-Off-The-Shelf Wrist Wearables to Estimate Stress on Students
Published on: June 16, 2018
Wrist-based wearable oscillometric blood pressure monitoring during Ramadan: a comparative study with standard
1Department of Electrical and Computer Engineering, King Abdulaziz University, Jeddah, Saudi Arabia.
Abstract:
The determination of blood pressure (BP) is crucial for the regulation of cardiovascular health status, particularly under the condition of physiological stress as seen during Ramadan fasting. Conventional BP measurement techniques are, accurate, intermittent and do not reflect the dynamic changes of fasting and rehydration. Wrist-based wearable oscillometric blood pressure monitoring devices, incorporating miniaturized inflatable cuffs, provide a portable and non-invasive approach for assessing BP outside traditional clinical settings. To assess the reliability of a consumer-grade wearable device (Huawei Watch D) for measuring systolic and diastolic blood pressure during Ramadan fasting, and to compare its performance with a validated electronic upper-arm monitor after pre-fast and post-fast conditions (Sahoor and If tar).
Methods:
A within subject, repeated-measures investigation was performed with 101 healthy fasting subjects (ages 22-70). Each participant recorded BP twice daily during Sahoor and If tar using both the Huawei Watch D and an upper-arm electronic monitor, both are FDA-approved. Statistical methods comprised paired t tests, as well as Pearson correlations and Bland-Altman plots. To evaluate prediction accuracy and discover subgroups, we also used machine learning approaches (linear regression and K-means clustering).
Results:
Compared with the electronic upper-arm monitor, the Huawei Watch D demonstrated minimal mean bias for both systolic and diastolic blood pressure (SBP: +0.9 to +1.4 mmHg; DBP: -0.6 to -0.2 mmHg). However, SBP exhibited wide limits of agreement (±18-21 mmHg), whereas DBP showed narrower dispersion (±11-13 mmHg). Machine-learning models achieved moderate prediction accuracy for DBP (MAE ≈ 4.3-6.1 mmHg, R² up to 0.74), while SBP prediction remained less accurate (MAE ≈ 6.5-8.9 mmHg, R² up to 0.64).However, SBP exhibited wide limits of agreement (±20 mmHg), limiting diagnostic interchangeability. DBP showed narrower error dispersion, supporting the use of wearable devices for trend monitoring and screening rather than clinical diagnosis.
Conclusions:
Smartwatch BP estimates are associated with standard devices, however, individual differences are considerable and therefore do not allow for diagnostic interchangeability. Wearables, such as the Huawei Watch D, provide the interesting data on the BP trend during fasting and rehydration, especially in diastolic pressure. These devices may assist with culturally appropriate monitoring and early identification of BP abnormalities however are in need of validation in a clinical context.
Related Concept Videos
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Assessment of blood pressure in brachial artery(two-step method)
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement