Related Experiment Video
Updated: May 19, 2026

In Vivo Imaging of Cx3cr1gfp/gfp Reporter Mice with Spectral-domain Optical Coherence Tomography and Scanning Laser Ophthalmoscopy
Published on: November 11, 2017
Short-Term Visual and Optical Performance of the ZEISS CT LUCIA 621P Intraocular Lens Across Different Power Ranges
Jiwon Jeong1,2, Younghee Kim2, Dong Wook Kim3
1Fatima Eye Clinic, Changwon, Republic of Korea.
Purpose:
To evaluate the short-term efficacy and optical performance of the ZEISS CT LUCIA 621P intraocular lens (IOL) in cataract patients using a diopter (D)-stratified approach.
Patients And Methods:
This retrospective study included 304 eyes of 216 patients who underwent phacoemulsification with CT LUCIA 621P implantation. Eyes were stratified into four groups by implanted IOL power: Group 1 (<20.0 D, n=49), Group 2 (20.0-21.5 D, n=137), Group 3 (22.0-23.5 D, n=60), and Group 4 (≥24.0 D, n=58). Primary outcomes were monocular and binocular photopic/mesopic defocus curves and internal higher-order aberrations (HOAs) at 1 and 3 months because these directly evaluate the optical quality and depth of focus delivered by the non-constant aspheric ZEISS Optic design-the core innovative feature of the CT LUCIA 621P IOL-across the full spectrum of implanted powers. Secondary outcomes provide complementary clinical efficacy data; uncorrected and corrected distance visual acuity (UDVA/CDVA), uncorrected intermediate (UIVA) and near (UNVA), refractive predictability, and contrast sensitivity (CS). Analyses used generalized estimating equations (GEE) with an exchangeable correlation structure.
Results:
All groups achieved excellent refractive predictability with convergence of spherical equivalent (SE) toward emmetropia (p=0.398). Mean monocular UDVA at 3 months was comparable across groups (p=0.498). Monocular and binocular UIVA and UNVA were also comparable (all p>0.23). Photopic and mesopic defocus curves showed similar depth of focus across groups (p>0.05 at most levels). Postoperative spherical aberration (SA) and coma remained low with no significant inter-group differences after generalized GEE adjustment (p>0.17). Photopic CS was comparable across IOL power groups at most spatial frequencies.
Conclusion:
The CT LUCIA 621P IOL demonstrated consistent refractive predictability, VA across distances, and optical quality across a wide range of implanted powers, supporting its reliable use in contemporary cataract surgery.
More Related Videos
06:55Scanning Light Scattering Profiler (SLPS) Based Methodology to Quantitatively Evaluate Forward and Backward Light Scattering from Intraocular Lenses
Published on: June 6, 2017
04:59Rotating the Intraocular Lens to Prevent Posterior Capsular Opacification in Cataract Surgeries
Published on: July 7, 2023
Related Concept Videos
Focusing of Light in the Eye
Imaging Biological Samples with Optical Microscopy
In optical microscopy, the specimen to be viewed is placed on a glass slide and clipped on the stage...