Two digital electrocardiographs and their computer-based analysis programs produce clinically significant differences
Peter Constable1, Tabitha Jones2, Christina Wilson2
1Department of Veterinary Clinical Medicine, College of Veterinary Medicine, University of Illinois Urbana-Champaign, Urbana, IL.
Veterinary and human electrocardiographs yield different canine ECG measurements, impacting reference intervals. Device filtering characteristics significantly influence results for endurance-trained dogs.
Area of Science:
- Veterinary Cardiology
- Biomedical Engineering
- Canine Physiology
Background:
- Electrocardiography (ECG) is crucial for assessing canine cardiac health.
- Establishing accurate reference intervals for ECG parameters in athletic dogs is essential for proper diagnosis.
- Differences in electrocardiograph (ECG) technology may influence measured values.
Purpose of the Study:
- To compare ECG measurements between veterinary and human electrocardiographs in endurance-trained dogs.
- To develop specific ECG reference intervals for this canine population.
- To identify how device characteristics affect ECG readings.
Main Methods:
- Simultaneous standard limb lead ECG recordings were obtained from 121 endurance-trained dogs using two digital electrocardiographs (veterinary and human).
- Computer-based ECG measurements were compared using statistical tests (Wilcoxon signed-rank, Deming regression) and Bland-Altman plots.
- Predefined acceptable median differences (effect sizes) were used to assess clinical significance.
Main Results:
- Significant differences were found in most global ECG indices and lead amplitudes between the two electrocardiographs.
- The veterinary ECG showed higher Q- and R-wave amplitudes (11-23%) in leads II and III.
- The veterinary ECG identified more dogs with potential ventricular enlargement and produced narrower reference intervals for most indices.
Conclusions:
- Veterinary and human electrocardiographs produce different ECG measurements in endurance-trained dogs.
- The low-pass filter characteristics of the veterinary ECG contribute to observed differences.
- Accurate canine ECG reference intervals are dependent on the specific electrocardiograph, its filtering, and analysis software.
More Related Videos
Related Concept Videos
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and the T...
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin to...
Dysrhythmias V: Evaluating Dysrhythmias
ECG Interpretation of Rhythms
Components of the Electrocardiogram
The primary components of a normal ECG waveform in Normal sinus rhythm(NSR) include the P wave, PR interval, QRS complex, ST segment, T wave, and occasionally a U wave.
ECG waveforms are divided by vertical and horizontal lines at standard intervals.
The horizontal axis measures time and rate, and the vertical axis measures amplitude or voltage. When...
Correlation between ECG and Cardiac Cycle
A cardiac action potential originates in the SA node and spreads throughout the atria and the AV node in approximately 0.03 seconds. This results in the P wave in an ECG and triggers atrial contraction. The action potential is then briefly slowed at the AV node, allowing the atria to contract and fill the ventricles with blood before...