The stability of diffusion measurements with simultaneous multi-slice imaging: a multi-scanner, multi-site,
Matthew Marzetti1,2,3, Hamied A Haroon4,5, William Lloyd4,5
1Department of Medical Physics, Leeds Teaching Hospitals NHS Trust, Leeds, United Kingdom.
Biomedical Physics & Engineering Express
|May 21, 2026
Summary
Simultaneous multi-slice (SMS) MRI acceleration can introduce bias in apparent diffusion coefficient (ADC) measurements, especially at higher factors. Low SMS acceleration is suitable for clinical use, maintaining ADC accuracy.
Area of Science:
- Magnetic Resonance Imaging (MRI)
- Quantitative Imaging
- Medical Physics
Background:
- Apparent diffusion coefficient (ADC) measurements are crucial for quantitative diffusion MRI.
- Simultaneous multi-slice (SMS) acceleration is increasingly used to reduce scan times.
- The impact of SMS acceleration on ADC accuracy across different MRI platforms is not fully understood.
Purpose of the Study:
- To evaluate the effect of SMS acceleration on ADC measurement accuracy and repeatability.
- To compare ADC measurements across multiple MRI platforms using a traceable diffusion phantom.
- To determine the optimal SMS acceleration factor for quantitative diffusion imaging.
Main Methods:
- A multi-center, multi-vendor test-retest study was performed using a CaliberMRI diffusion phantom.
- Diffusion-weighted images were acquired with SMS acceleration factors of 0, 2, 3, and 4.
- ADC values were compared to NIST reference standards, and signal-to-noise ratio (SNR) was calculated.
Main Results:
- Scanners overestimated ADC by an average of 4%.
- SMS factors 3 and 4 introduced significant systematic biases (4.62% and 9.22% overestimation, respectively).
- Low SMS acceleration (factor 2) did not significantly bias ADC measurements, and repeatability remained high despite reduced SNR.
Conclusions:
- Higher SMS acceleration factors (≥3) can introduce bias and degrade image quality in ADC measurements.
- Low SMS acceleration (factor 2) is suitable for reducing scan times without compromising ADC accuracy.
- Caution is needed when using higher SMS factors for absolute ADC quantification across platforms.