Related Experiment Video
Updated: May 28, 2026

09:57
Ultrafast Time-resolved Near-IR Stimulated Raman Measurements of Functional π-conjugate Systems
Published on: February 10, 2020
High-Performance SiPM Detection Module for Ultra-Fast Time-Resolved Measurements.
Gennaro Fratta1, Piergiorgio Daniele1, Ivan Labanca1
1Department of Electronics, Information and Bioengineering (DEIB) of Politecnico di Milano, 20133 Milan, Italy.
Sensors (Basel, Switzerland)
|May 27, 2026
Summary
A new Time-Correlated Single-Photon Counting (TCSPC) method overcomes pile-up distortion, enabling faster, distortion-free light signal analysis. This advancement utilizes a Silicon Photomultiplier detector for enhanced precision in biomedical applications.
Area of Science:
- Biomedical optics
- Photonics
- Light-matter interaction analysis
Background:
- Non-invasive light-matter interaction analysis is rapidly advancing in biomedical and life sciences, driven by low-intensity light detection.
- Single-photon detection techniques, particularly Time-Correlated Single-Photon Counting (TCSPC), are crucial for analyzing complex photonic applications.
- Conventional TCSPC is limited by pile-up distortion, restricting acquisition speed and signal fidelity.
Purpose of the Study:
- To introduce a novel TCSPC acquisition methodology that overcomes limitations of conventional implementations.
- To develop a single-photon detection module for improved TCSPC performance.
- To enable distortion-free reconstruction of light profiles independent of photodetector dead time and excitation intensity.
Main Methods:
- Development of a novel TCSPC acquisition methodology.
- Implementation of a single-photon detection module using a Silicon Photomultiplier.
- Performance evaluation through fluorescence measurements using the constraint-free TCSPC methodology.
Main Results:
- The developed Silicon Photomultiplier module provides 750 ps FWHM output pulses with a 33.5 ps RMS Instrument Response Function (IRF).
- The constraint-free TCSPC methodology achieved a photon count rate up to 166% of the excitation frequency.
- Minimal lifetime estimation error of -1.46% was recorded, demonstrating high accuracy.
Conclusions:
- The novel TCSPC methodology and Silicon Photomultiplier module significantly enhance the capabilities of time-resolved optical signal analysis.
- This advancement overcomes critical limitations of traditional TCSPC, enabling higher acquisition speeds and improved accuracy.
- The developed system holds promise for transforming biomedical and life science research reliant on precise light detection.