Imaging Metals in Brain Tissue by Laser Ablation - Inductively Coupled Plasma - Mass Spectrometry (LA-ICP-MS)
Published on: January 22, 2017
Aliasing Artifacts in Multielemental LA-ICP-TOFMS Mapping: An Old Problem Revisited
Filip Cernatič1, Kristina Mervič2, Lukas Schlatt3
1Department of Analytical Chemistry, National Institute of Chemistry, Hajdrihova ulica 19, SI-1001 Ljubljana, Slovenia.
Signal aliasing in laser ablation inductively coupled plasma time-of-flight mass spectrometry (LA-ICP-TOFMS) is unavoidable. Optimizing acquisition time balances signal spread and data redundancy to minimize these artifacts.
Area of Science:
- Analytical Chemistry
- Mass Spectrometry
- Geochemistry
Background:
- Unsynchronized laser pulses and data acquisition in quadrupole LA-ICP-MS cause signal aliasing, leading to inaccurate quantitation and degraded image quality.
- Signal aliasing manifests as low- to medium-frequency alterations in time-averaged signals, not reflecting true analyte concentrations.
Purpose of the Study:
- To investigate and address signal aliasing in LA-ICP-TOFMS.
- To determine optimal acquisition time settings in TOF instruments to mitigate aliasing artifacts.
Main Methods:
- Theoretical simulations of signal acquisition in TOF mass analyzers.
- Experimental mapping of standard reference materials using LA-ICP-TOFMS.
Main Results:
- Exact synchronization between laser pulses and data acquisition is impossible in LA-ICP-TOFMS due to constrained acquisition times.
- Signal aliasing is an inherent issue in LA-ICP-TOFMS, even with acquisition times near the SPR.
- Optimal acquisition time involves a compromise to minimize signal spread while avoiding excessive redundant data.
Conclusions:
- Minimizing signal aliasing in LA-ICP-TOFMS requires careful selection of acquisition time.
- A balance must be struck between reducing signal spread and managing data volume for accurate quantitative analysis and imaging.