Gentle limb fixation for a faster reliable heart rate obtained with pulse oximeter at birth
Mirthe M Bloemhof1, Janneke Dekker1, Sanne Ma van den Berg1
1Willem-Alexander Children's Hospital, Department of Paediatrics, Division of Neonatology, Leiden University Medical Centre, the Netherlands.
Aim:
Pulse oximetry is widely used to monitor heart rate (HR) and peripheral oxygen saturation during neonatal resuscitation. As HR is a key parameter guiding clinical decision-making during initial assessment and stabilization, timely and accurate HR readings are critical. However, it can take minutes to achieve a reliable reading directly after birth. This study aimed to determine whether infant limb movement and gentle fixation after pulse-oximeter application influence the time to obtain the first reliable HR.
Methods:
Video recordings of neonatal resuscitations with simultaneous pulse oximetry readings were retrospectively analyzed for neonates ≥24 weeks' gestation. Neonates were categorized into three groups based on the status of the right limb during the first 10 s after sensor application: (1) movement, (2) no movement, or (3) gentle fixation by a caregiver. Time to the first reliable HRPO was determined using a custom-developed, ECG-validated algorithm applied to the photoplethysmogram.
Results:
A total of 178 neonates were included, with a median gestational age of 30 + 4 weeks (28 + 4-33 + 6), with no significant differences between groups. Median time to achieve a reliable HRPO was significantly shorter in the "no movement" (20.3 [16.1-34.2] s) and "fixation" (17.9 [14.9-25.4] s) groups as compared to the "movement" group (46.2 [23.5-75.1] s) (p < 0.001). The percentage of neonates who achieved a reliable HRPO within 30 s after application was significantly higher after limb fixation (80.5%) versus no fixation (49.6%) (p < 0.001).
Conclusion:
Pulse oximetry does not seem motion artefact resistant when the first HR is obtained. Gentle limb fixation after sensor application may facilitate more timely clinical assessment.
More Related Videos
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Special considerations while measuring pulse
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...