Related Experiment Video
Updated: Jun 13, 2026

Split Point Analysis and Uncertainty Quantification of Thermal-Optical Organic/Elemental Carbon Measurements
Published on: September 7, 2019
Comprehensive Comparison of ISO-GUM and Monte Carlo Simulation for Uncertainty Assessment in HPLC Statin
Hamza Belmir1, Hasnaa Haidara2, Rajae Aitlhaj3
1Laboratory of Analytical and Molecular Chemistry, Faculty of Sciences Ben M'Sick, Hassan II University, Morocco.
Background:
Reliable measurement uncertainty estimation is essential for building trust in analytical results and ensuring they meet regulatory expectations. In HPLC chromatography, the chosen approach for uncertainty estimation can strongly influence both the results and their interpretation.
Objective:
This study compares two widely used strategies, the ISO-GUM method and Monte Carlo simulation applied with either single-point or multi-point calibration, for the simultaneous quantification of four statins at the National Laboratory for Drug Control (LNCM).
Methods:
In the single-point approach, all sources of uncertainty were combined into one global model, while the multi-point approach evaluated variability across different calibration levels. Both strategies were applied within the ISO-GUM framework and through Monte Carlo simulation, allowing a side-by-side comparison of the resulting uncertainty budgets.
Results:
The single-point method tended to underrepresent the variability present in actual measurements. In contrast, the multi-point method, especially when paired with Monte Carlo simulation, delivered uncertainty estimates that better reflected real-world analytical conditions. The ISO-GUM framework combined with multi-point calibration also proved to be a robust and practical option for routine use.
Conclusions:
Multi-point calibration, whether combined with Monte Carlo simulation or the ISO-GUM method, offers a more representative and reliable way to assess uncertainty in HPLC statin analysis.
More Related Videos
10:17High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry
Published on: April 23, 2019
08:53Quantitative and Qualitative Method for Sphingomyelin by LC-MS Using Two Stable Isotopically Labeled Sphingomyelin Species
Published on: May 7, 2018
Related Concept Videos
Uncertainty: Overview
Propagation of Uncertainty from Systematic Error
Data Validation
Key parameters for method validation include:
In Vitro Drug Dissolution: Compendial Testing Models I
In Vitro Drug Dissolution: Compendial Testing Models II
Pharmacokinetic Models: Comparison and Selection Criterion
Physiological models take a detailed approach by considering specific molecular processes. They can predict drug distribution, metabolism, and elimination changes, providing a comprehensive understanding of how drugs interact with the body.