Related Experiment Video
Updated: Jun 16, 2026

Measuring the Carotid to Femoral Pulse Wave Velocity (Cf-PWV) to Evaluate Arterial Stiffness
Published on: May 3, 2018
Accuracy and precision of continuous, non-invasive blood pressure devices: a systematic review and meta-analyses
Stefanie Lischker1, Gunther J Pestel1, Irene Schmidtmann2
1Department of Anaesthesiology, University Medical Center of the Johannes Gutenberg-University, Mainz, Germany.
Introduction:
Reliable continuous non-invasive blood pressure monitoring technology should support clinicians to maintain blood pressure within appropriate limits peri-operatively. Previous studies have shown conflicting results regarding measurement accuracy.
Methods:
This systematic review and six meta-analyses evaluated the accuracy and precision of non-invasive continuous arterial blood pressure measurement technologies (volume clamp technology, applanation tonometry and photoplethysmography) compared with invasive arterial measurement in adults. We analysed pooled bias (mean difference between non-invasive and invasive measurement), standard deviation and limits of agreement. In each meta-analysis, mean arterial pressure (MAP), systolic blood pressure (SBP) and diastolic blood pressure (DBP) were analysed separately. Acceptability was defined according to Association for the Advancement of Medical Instrumentation® (AAMI) criteria.
Results:
Sixty-five studies were included: two meta-analyses for volume clamp technology (ClearSight™, CNAP™); three meta-analyses for applanation tonometry (T-Line™, Vasotrac™ and N-CAT™); and one meta-analysis for photoplethysmography technology. For volume clamp technologies, none met AAMI criteria for bias and SD in relation to MAP, SBP and DBP. T-Line, Vasotrac and N-CAT met the AAMI criteria for bias for MAP (2.53 mmHg, 0.36 mmHg, -0.37 mmHg), SBP (-0.07 mmHg, 1.20 mmHg, -3.68 mmHg) and DBP (3.09 mmHg, 1.09 mmHg, 0.84 mmHg), respectively. In terms of standard deviation, only N-CAT (MAP 6.97 mmHg, SBP 7.14, mmHg, DBP 6.55 mmHg) met AAMI criteria for all blood pressure parameters. Photoplethysmography technology met AAMI criteria for bias (MAP -0.09 mmHg, SBP 0.49 mmHg, DBP -0.26 mmHg) and standard deviation (MAP 6.86 mmHg, DBP 6.38 mmHg).
Discussion:
Overall, the AAMI criteria for bias and SD were not fulfilled for most of the devices tested and wide limits of agreement are of concern. One applanation tonometry device (N-CAT) shows promise. Clinicians should be aware of the lack of interchangeability with invasive blood pressure management.
Related Concept Videos
Errors occurring during blood pressure monitoring
Several factors...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(two-step method)
Pre-Procedural Guidelines for Assessing Blood Pressure
