Influence of airway pressure and fibrosis on lung surface shear wave elastography
Yuki Kawakami1, Kensuke Nakamura2, Hironobu Yasui3
1Laboratory of Veterinary Internal Medicine, Department of Veterinary Clinical Sciences, Graduate School of Veterinary Medicine, Hokkaido University, N18 W9, Sapporo, Hokkaido, 060-0818, Japan.
Purpose:
Lung ultrasound is widely used in clinical practice, but B-mode findings often lack disease specificity. Shear wave elastography (SWE) enables quantitative assessment of tissue elasticity and may help overcome this limitation. This study aimed to evaluate whether SWE can be used to detect lung surface changes associated with airway pressure and pulmonary fibrosis.
Methods:
Ex vivo porcine lungs (n = 3) were used to examine the effects of airway pressure. Airway pressure was adjusted to 10, 20, and 30 cmH2O, and shear wave speed (SWS) was measured on the lung surface. Pulmonary fibrosis was induced in Sprague-Dawley rats (n = 13) by means of right thoracic irradiation. After 26 weeks, SWE was performed on excised lungs, and histopathological analysis of fibrosis and air proportions was conducted using QuPath-based image analysis. Statistical analyses included Spearman's correlation, Wilcoxon signed-rank test, and multiple regression with isometric log-ratio (ILR) transformation.
Results:
In porcine lungs, SWS increased with airway pressure and showed a strong positive correlation. In the radiation-induced fibrosis model, irradiated lungs had significantly higher SWS than non-irradiated lungs. Histopathology confirmed increased fibrosis and reduced air without inflammation. Multiple regression showed a significant association between fibrosis and SWS.
Conclusion:
Lung surface SWE detected changes related to airway pressure and pulmonary fibrosis. SWS increased with both, although other factors, such as lung reduction in air content and tensile stress, can also affect measurements.
More Related Videos
05:56Implementation of Non-invasive Point of Care Transient Elastography for Evaluation of Liver Disease in Pediatric Populations with Cystic Fibrosis
Published on: August 29, 2025
07:13Application of Ultrasound and Shear Wave Elastography Imaging in a Rat Model of NAFLD/NASH
Published on: April 20, 2021
Related Concept Videos
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):
Pressure Relationships in Thoracic Cavity
Breathing Mechanisms
Both intra-alveolar and intrapleural pressures rely on specific lung properties. The ability to breathe—allowing air to enter the lungs during...