Related Experiment Video
Updated: Jun 18, 2026

Mouse Body Temperature Measurement Using Infrared Thermometer During Passive Systemic Anaphylaxis and Food Allergy Evaluation
Published on: September 14, 2018
Comparing the Accuracy of Patient-Perceived Fever, Manual Fever Checking, and Non-Contact Frontal Infrared
Moojan Shabani-Kakroudi1, Mohammad Jalili1,2, Ali Pasha Meysamie3
1Emergency Medicine Department, Imam Khomeini Hospital Complex (IKHC), Tehran University of Medical Science (TUMS), School of Medicine, Tehran, Iran.
Introduction:
Non-contact frontal infrared thermometers (NCFITs) have gained popularity following the COVID-19 pandemic, but their accuracy remains debated. This study compared the accuracy of three index tests of patient-perceived fever (PPF), manual fever checking (MFC) by emergency nurses, and NCFIT against tympanic thermometer (TT).
Methods:
This single center, diagnostic accuracy study evaluated adult patients in the triage area of the emergency department (ED) between May and September 2023. Five emergency nurses (ENs) examined patients using convenience sampling with index tests, followed by TT measurement for fever detection as standard test (T≥38°C). The sensitivity, specificity, positive and negative likelihood ratios (PLR and NLR) and accuracy, of PPF, MFC, and NCFIT in detecting the fever were measured and reported with 95% confidence interval (CI). The inter-rater reliability of ENs for MFC among 10 patients (outside the study) was 100%.
Results:
We included 279 patients with the mean age of 52.2 ± 16.98 years (51.97% male). TT identified 147 as febrile and 132 as non-febrile. The most sensitive index test was MFC, achieving 100% (95%CI: 97.5-100.0) sensitivity, 0 NLR, and 8.25 (95%CI: 5.7-12.1) PLR, with moderate accuracy (93.9% (95%CI: 89.9-96.0)). NCFIT demonstrated the lowest sensitivity (85.71% (95%CI: 78.9-90.9)) but the highest specificity (98.48% (95%CI: 93.6-99.5)) and accuracy (97.5% (95%CI: 93.1-99.8)). When the fever threshold for NCFIT was lowered to 37.5°C, sensitivity increased to 99% (95%CI: 97.4-99.9), NLR decreased to 0.01 (95%CI: 0-0.6), and PLR rose to 49.5 (95%CI: 20.8-90.2), while specificity remained at 98% (95%CI: 94-99.6).
Conclusion:
MFC/PPF could rule out fever but required NCFIT/TT to rule it in. However, NCFIT achieved the highest NLR and PLR when its fever threshold was set at 37.5 °C.
More Related Videos
Related Concept Videos
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's forehead...
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Assessing Body Temperature - Rectal
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.
Step 2: Position the patient in a side-lying position to better visualize the rectal...