Dual-energy CT pulmonary angiography using dual-flow mixture bolus tracking: potential for contrast medium and
Mingjue Jian1,2, Lisheng Huang1, Chenxi Zhang1,2
1Department of Medical Imaging Center, Nanfang Hospital, Southern Medical University, Guangzhou, China.
Objectives:
To assess the feasibility and image quality of a bolus tracking protocol with dual-flow mixture for contrast medium volume and radiation dose reduction in dual-energy CT pulmonary angiography (CTPA).
Materials And Methods:
Patients with suspected pulmonary embolism (PE) were prospectively included and randomly divided into the dual low-dose or routine CTPA groups. Virtual monoenergetic imaging results at 40 keV for the dual low-dose CTPA group were compared to 100-kV polyenergetic CT images from the routine CTPA group. Attenuation, noise, contrast-to-noise ratio, and figure of merit were determined in multiple pulmonary arteries. Qualitative image quality and PE detection were independently rated by two radiologists.
Results:
The dual low-dose CTPA group showed a significantly higher attenuation in 40-keV virtual monoenergetic imaging (1027.0 ± 287.6 HU vs 391.8 ± 109.0 HU; p < 0.001) with a comparable median contrast-to-noise ratio (33.6 vs 41.0; p = 0.115) and superior figure of merit (711.4 vs 461.5; p = 0.006) in the pulmonary trunk in relation to that observed in the routine CTPA group. Superior vena cava artifacts were reduced (both p ≤ 0.002), while pulmonary branch visualization was preserved (p = 0.660 and 0.763). Effective radiation dose was lower in the dual low-dose CTPA group (1.61 mSv vs 3.62 mSv; p < 0.001), while PE diagnostic performance did not differ significantly between the two protocols.
Conclusion:
The dual-flow mixture bolus-tracking protocol showed technical feasibility for dual-energy CTPA, achieving substantial reductions in radiation exposure and contrast medium volume while maintaining diagnostically acceptable image quality.
Critical Relevance Statement:
The dual-flow mixture bolus-tracking protocol with dual-energy CT maintained feasible image quality for PE assessment, facilitating contrast medium volume and radiation exposure reduction.
Trial Registration:
NFEC, NFEC-2025-021, approved on January 10, 2025.
Key Points:
Radiation exposure and substantial contrast medium doses during CTPA increased patient risks. Virtual monoenergetic imaging at 40 keV derived from dual-energy CT enhanced pulmonary vascular visualization. The dual-flow mixture technique with dual-energy CT reduced radiation dose and contrast medium volume.
More Related Videos
09:49Dual-phase Cone-beam Computed Tomography to See, Reach, and Treat Hepatocellular Carcinoma during Drug-eluting Beads Transarterial Chemo-embolization
Published on: December 2, 2013
09:32Time-Resolved, Dynamic Computed Tomography Angiography for Characterization of Aortic Endoleaks and Treatment Guidance via 2D-3D Fusion-Imaging
Published on: December 9, 2021
Related Concept Videos
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies for Cardiovascular System V: CT
Radiological Investigation I: X-ray and CT