Related Experiment Video
Updated: Jun 23, 2026

Assessment of Thermal Damage from Robot-Drilled Craniotomy for Cranial Window Surgery in Mice
Published on: November 11, 2022
Drill choice as a modifiable factor in pediatric sEEG safety
Mahalia Dalmage1, Sunny Abdelmageed2,3, Lucinda Chiu2,4
1Division of Biological Sciences, University of Chicago Pritzker School of Medicine, Chicago, IL, USA.
A new neurosurgery-tailored drill (Phasor®) demonstrated consistent and time-efficient burr hole creation for stereoelectroencephalography (sEEG) procedures. This drill offers a safe alternative to repurposed orthopedic drills, with no complications noted in the case series.
Area of Science:
- Neurosurgery
- Medical Devices
- Surgical Techniques
Background:
- Stereoelectroencephalography (sEEG) requires precise burr hole creation for electrode placement.
- Current sEEG practices often repurpose orthopedic drills, which may lack optimal characteristics.
- Key drilling parameters for sEEG include haptic feedback, ease of use, and efficiency.
Purpose of the Study:
- To evaluate a novel neurosurgery-tailored drill (Phasor®) for sEEG procedures.
- To compare the performance of the Phasor® drill against a repurposed orthopedic drill (Stryker®).
- To assess drilling characteristics such as time efficiency, haptic feedback, and safety.
Main Methods:
- Retrospective chart review of patients undergoing sEEG from 2022-2023.
- Patients were grouped based on the drill used: Phasor® (neurosurgery-tailored) or Stryker® (orthopedic).
- Data collected included number of electrodes placed, operative duration, and complications.
Main Results:
- The Phasor® drill group (12 patients) showed consistent burr hole creation time (<15s) and noted tactile bony feedback.
- The Stryker® drill group (5 patients) exhibited variable drill times and one complication (skin burn leading to infection).
- Mean operative duration was comparable, but the Phasor® drill offered subjective improvements in consistency and safety.
Conclusions:
- The Phasor® drill is a lightweight, reliable, and efficient option for sEEG burr hole creation.
- It provides consistent drilling performance and improved safety compared to repurposed orthopedic drills.
- The Phasor® drill is a viable alternative for neurosurgical procedures requiring precise burr hole creation.