Aortic dissection false lumen treatment with novel shape memory polymer devices
Related Experiment Video
Updated: Jun 25, 2026

Novel and Innovative Hybrid Technique for Type A Aortic Dissection
Published on: March 28, 2025
Dai Yamanouchi1,2, Yusuke Sakurai2, Masaru Nemoto2
1Division of Vascular Surgery, Department of Surgery, University of Wisconsin, Madison, WI.
Abstract:
In this case series, innovative shape memory polymer (SMP) devices were used to treat the aortic dissection false lumen (FL) of five patients. The open-cell, porous structure of expanded SMP contributes to its radiolucency, low radial force, and conformability. The devices were designed to address the unique needs of treating the FL: sufficiently large volume, detachable and stable to minimize migration risk, segmented for flexibility, and with a range of dimensions. All five cases were technically successful. A decrease in total aortic diameter, true lumen restoration, and FL reduction were observed through 6- to 12-month follow-up. The mean changes in maximum aorta diameter from preoperative measurements were -5.8 ± 6.9 mm (30 days, n = 5), -12.0 ± 9.4 mm (6 months, n = 4), and -9.3 ± 0.8 mm (12 months, n = 2). No SMP device/procedure-related complications occurred.
