Patient Directed Recording of a Bipolar Three-Lead Electrocardiogram using a Smartwatch with ECG Function
Published on: December 11, 2019
Smartwatch-Based Heart Rate Detection in Atrial, Ventricular and Pulseless Arrhythmias
Samual Turnbull1, Anunay Gupta1, Kartheek Garikapati1
1Department of Cardiology, Westmead Hospital, Westmead Applied Research Centre, University of Sydney, Westmead, NSW, Australia.
Smartwatches show good heart rate accuracy for regular rhythms but struggle with detecting arrhythmia terminations and ventricular arrhythmias. Use caution when relying on wearables for cardiac arrhythmia monitoring.
Area of Science:
- Cardiology
- Biomedical Engineering
- Wearable Technology
Background:
- Smartwatches offer non-invasive heart rate (HR) monitoring for atrial fibrillation screening.
- Accuracy of HR detection during other arrhythmias needs further study.
Purpose of the Study:
- Assess HR measurement accuracy of Apple Watch (AW) and Fitbit (FB) during induced and paced arrhythmias.
- Evaluate device performance at onset, during, and termination of arrhythmias.
Main Methods:
- Intraprocedural fitting of AW and FB on patients undergoing electrophysiology studies.
- Recording HRs during various paced/induced arrhythmias; confirmed with intracardiac electrograms and 12-lead ECG.
- Assessed accuracy as HR within ±10% of the true rate.
Main Results:
- Analyzed 322 cardiac rhythms from 32 patients.
- AW and FB showed reasonable HR detection during sinus rhythm (100% vs 87%) and supraventricular tachycardia (SVT) (89% vs 86%).
- Both devices performed poorly during SVT termination (62% vs 41%) and ventricular tachycardia (41% vs 16%); failed to detect HR changes in ventricular fibrillation or asystole.
Conclusions:
- Apple Watch and Fitbit offer reasonable HR accuracy for regular supraventricular rhythms.
- Devices struggle with arrhythmia terminations and ventricular arrhythmias.
- Caution is advised when using HR trends from wearables to assess cardiac arrhythmias.
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Holter Monitor: 24-Hour Monitoring
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Dysrhythmias V: Evaluating Dysrhythmias
Special considerations while measuring pulse
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation