Related Experiment Video
Updated: Jun 30, 2026

Robotic Ablation of Atrial Fibrillation
Published on: May 29, 2015
Multicatheter Comparison of Very High-Power Short-Duration Radiofrequency Ablation Lesions
Samual Turnbull1,2, Michael A Barry1,2, Timothy G Campbell1,2
1Department of Cardiology, Westmead Hospital, Westmead, New South Wales, Australia.
Catheter design impacts very high-power short-duration (vHPSD) radiofrequency ablation (RFA) lesion characteristics. The QDOT catheter produced deeper lesions than the TactiFlex, while the SmartTouch catheter yielded the widest lesions and highest steam-pop risk.
Area of Science:
- Electrophysiology
- Medical Devices
- Cardiovascular Interventions
Background:
- Very high-power short-duration (vHPSD) radiofrequency ablation (RFA) is an alternative strategy for pulmonary vein isolation.
- Rapid temperature rises during RFA can lead to complications.
- The QDOT Micro catheter features automatic power and irrigation adjustments based on temperature detection.
Purpose of the Study:
- To compare the impact of different ablation electrode designs on vHPSD lesions.
- To evaluate lesion characteristics generated by three distinct RFA catheter designs.
Main Methods:
- vHPSD RFA was performed in a gel tank model using QDOT, SmartTouch, and TactiFlex SE catheters.
- Ablations were delivered at varying power settings (60-90 watts) and irrigation concentrations (0.9% and 0.45% half-normal saline).
- Lesions were characterized based on depth, diameter, and temperature to assess steam-pop risk.
Main Results:
- The SmartTouch catheter produced the deepest (2.37 mm) and widest (6.04 mm) lesions.
- The QDOT catheter generated deeper lesions (1.93 mm) compared to the TactiFlex (1.88 mm).
- The TactiFlex catheter exhibited the lowest risk of steam pops.
Conclusions:
- Catheter electrode design influences vHPSD lesion dimensions and temperature.
- The QDOT catheter achieved greater lesion dimensions than the TactiFlex, indicating differences in cooling profiles.
- Using the SmartTouch catheter or 0.45% half-normal saline may increase the risk of steam pops.