A Nondiagnostic 99mTc-PYP Scan with Absent Skeletal Uptake.
Hiroyuki Tokue1, Azusa Tokue1, Yoshito Tsushima1
1Department of Diagnostic and Interventional Radiology, Gunma University Hospital, 3-39-22 Showa-Machi, Maebashi 371-8511, Gunma, Japan.
Technetium-99m pyrophosphate (PYP) scintigraphy for cardiac amyloidosis can be nondiagnostic if radiotracer biodistribution is abnormal. Gastric and thyroid uptake with absent skeletal activity suggests an error, not a negative result.
Area of Science:
- Nuclear Medicine
- Cardiology
- Radiopharmacy
Background:
- 99mTc-PYP scintigraphy is a key noninvasive tool for diagnosing transthyretin cardiac amyloidosis.
- Accurate interpretation requires assessing both myocardial uptake and systemic radiotracer biodistribution.
- Abnormal biodistribution can mimic or obscure true findings.
Purpose of the Study:
- To report a case of 99mTc-PYP scintigraphy with abnormal biodistribution mimicking free pertechnetate.
- To highlight the importance of recognizing incorrect biodistribution for accurate diagnosis.
- To emphasize that such scans should be considered nondiagnostic, not negative.
Main Methods:
- A case study of a 75-year-old woman with suspected cardiac amyloidosis.
- Initial 99mTc-PYP scintigraphy showed unexpected gastric and thyroid uptake with absent skeletal activity.
- SPECT/CT was used for detailed imaging. A repeat scan was performed 1.5 months later.
Main Results:
- The initial scan demonstrated a biodistribution pattern consistent with free 99mTc-pertechnetate, not 99mTc-PYP.
- No definite myocardial uptake was observed in either scan.
- The repeat scan showed expected skeletal uptake, confirming the initial scan's error.
Conclusions:
- Abnormal 99mTc-PYP biodistribution (gastric/thyroid uptake, absent skeletal activity) indicates a preparation or administration error.
- Such findings render the scan nondiagnostic for cardiac amyloidosis.
- Clinicians should be aware of this potential pitfall and consider repeat imaging if necessary.
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...