Development of remote radar based vital sign acquisition for emergency department patient triage
Yossi Shaya1,2, Gil Levy3,4, Akiva Esterson1,2
1Department of Emergency Medicine, Beilinson Hospital, Rabin Medical Center, Petah Tikva, Israel.
Introduction:
To enable remote acquisition of patients' vital signs, ELTA Systems Ltd. developed a Radar remote vital signs acquisition system (TAMAR). The system identifies chest and neck respiratiory movements and pulse wave movements, enabling quantification of respiratory and pulse rates. The validation process of the TAMAR system is reported.
Methods:
A multicenter, prospective observational study was performed, enrolling low acuity patients who presented to 3 Emergency Departments. Patients provided consent and had their pulse and respiratory rates simultaneously measured by the TAMAR radar system and a traditional monitor (Mindray BeneVision N15). Vital signs were recorded continuously by both systems for 150 s, and later compared using Bland-Altman statistical analysis.
Results:
Between May and September 2021 we enrolled 315 patients. Measurements were calculated in 181/259 (70%) patients for Heart Rate (HR), and in 191/252 patients (76%) for Respiratory Rate (RR). For RR, the upper level of agreement (LOA) was 1.0, and the lower LOA was 1.6 respirations per minute. For HR, the upper level of agreement was 4.7, and the lower LOA was 3.5 beats per minute. All of these were well below the predetermined acceptable LOA of +/- 10 percent of the reference monitor rate.
Discussion:
The vital signs acquired by the TAMAR system, based on 90-second average measurements, are statistically adequate compared to traditional monitors, suggesting that TAMAR may replace traditional monitors for patients' triage.
More Related Videos
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
05:01A Detailed Protocol for Physiological Parameters Acquisition and Analysis in Neurosurgical Critical Patients
Published on: October 17, 2017
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation
Assessment of radial pulse
The radial pulse, located at the wrist, is often the preferred site for assessing peripheral pulse because of its accessibility and dependability. The process of determining the radial pulse involves several steps:
Introduction to Vital Signs
Vital signs help healthcare professionals assess an individual's well-being and detect any functional changes or...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.