Integration of PRAETORIAN Score, defibrillation testing and follow-up outcomes for the assessment of conversion failure in patients with subcutaneous implantable cardioverter defibrillator: Results from the iSUSI registry
Related Experiment Video
Updated: Jun 28, 2026

A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
Marco Schiavone1, Alessio Gasperetti2, Roland Tilz3
1Department of Clinical Electrophysiology and Cardiac Pacing, Centro Cardiologico Monzino IRCCS, Milan, Italy.
Background:
Although reports of good clinical outcomes after avoiding defibrillation testing (DFT) during subcutaneous implantable cardioverter defibrillator (S-ICD) implantation have increased over recent years, current guidelines still recommend performing DFT at the time of implant.
Objective:
We aimed to identify predictors of conversion failure in a real-world S-ICD population and assess whether PRAETORIAN score performance is improved by additional clinical variables.
Methods:
We analyzed 1063 patients from the international subcutaneous implantable cardioverter defibrillator (iSUSI) registry with available imaging and complete follow-up. The primary endpoint was a composite of DFT failure and ineffective appropriate shocks during follow-up. PRAETORIAN score and individual components were assessed, and receiver operating characteristic curve analysis with DeLong's test was used to evaluate model performance.
Results:
Patients were classified into low (76.6%), intermediate (16.3%), and high-risk (7.2%) PRAETORIAN score categories. Among 748 patients who underwent DFT, 64 (8.6%) experienced DFT failure; 17 ineffective shocks occurred during follow-up. Primary outcome occurred in 4.5% of low, 6.7% of intermediate-, and 16.8% of high-risk patients. PRAETORIAN score step 1, anterior generator position at PRAETORIAN score step 2 (odds ratio 2.3 [1.2-4.2], P = .006), high body mass index (BMI) and post-shock impedance, independently predicted conversion failure. Overall PRAETORIAN score (area under the curve [AUC] 0.71) outperformed individual steps (step 1-3 AUCs: 0.68-0.70; P > .05 for all comparisons). A multivariable model combining PRAETORIAN score, BMI, and impedance significantly improved predictive accuracy (AUC 0.78 vs 0.71; P = .0078).
Conclusion:
PRAETORIAN score, BMI, and impedance are independent predictors of defibrillation failure. Their combined use may improve risk stratification and may help guide clinical decision-making, especially when DFT is omitted.