A Method for Quantifying Upper Limb Performance in Daily Life Using Accelerometers
Published on: April 21, 2017
Benchtop validation of a low-cost wearable triaxial accelerometer
Carine Freitas E Silva1,2, Iara Dos Santos Leal1, Paulo Roberto Varjal de Melo3
1Master's and Doctoral Programs in Rehabilitation and Functional Performance, University of Pernambuco, Petrolina, Brazil.
Biomedical Physics & Engineering Express
|July 6, 2026
Summary
A new low-cost triaxial accelerometer shows strong agreement with a commercial reference sensor, validating its use for detecting fall events in older adults through motion monitoring.
Area of Science:
- Biomedical Engineering
- Sensor Technology
- Gerontology
Background:
- Wearable inertial sensors are crucial for detecting fall-related events in older adults.
- Rigorous validation of low-cost devices is essential to ensure measurement fidelity and stability.
Purpose of the Study:
- To evaluate the concurrent validity of a new, low-cost triaxial accelerometer.
- To compare its linear acceleration measurements against a commercial reference sensor (RS).
Main Methods:
- A methodological validation study involving 49 sensor under test (SUT) units and a Physilog®5 RS.
- Simultaneous collection of triaxial linear acceleration data across three axes with repeated trials.
- Analysis using bias, root mean squared error (RMSE), mean absolute error (MAE), waveform comparison, and Bland-Altman analysis.
Main Results:
- The SUT showed minimal bias on X and Y axes, with slightly higher errors on the Z-axis.
- RMSE and MAE were similar for X and Y axes, with moderate increases on the Z-axis.
- Strong waveform overlap and near-unity correlations indicated high signal correspondence; Bland-Altman showed minimal bias and narrow limits of agreement for X and Y axes.
Conclusions:
- The low-cost accelerometer (~$60 USD) demonstrates strong agreement with a high-cost reference device (> $1,000 USD).
- The device is validated for three-dimensional linear acceleration measurement under controlled conditions.
- This accelerometer offers a promising, cost-effective solution for scalable motion monitoring and fall detection in older adults.