In vivo Structural Assessments of Ocular Disease in Rodent Models using Optical Coherence Tomography
Published on: July 24, 2020
Macular Layer Thickness with Spectral-Domain and High-Resolution OCT
Sung-Uk Baek1,2, Glen P Sharpe1, Lesya M Shuba1
1Department of Ophthalmology and Visual Sciences, Dalhousie University, Halifax, Nova Scotia, Canada.
Ophthalmology Science
|July 12, 2026
Summary
High-resolution OCT (HR-OCT) and standard spectral-domain OCT (SD-OCT) show small but consistent differences in retinal layer thickness measurements. These variations may impact monitoring disease progression when switching between OCT devices.
Area of Science:
- Ophthalmology
- Medical Imaging
- Retinal Imaging
Background:
- High-resolution optical coherence tomography (HR-OCT) offers improved axial resolution over standard spectral-domain OCT (SD-OCT).
- Accurate segmentation of retinal layers is crucial for monitoring diseases like glaucoma.
- Understanding inter-device differences in OCT measurements is vital for longitudinal patient follow-up.
Purpose of the Study:
- To compare retinal layer thickness measurements between HR-OCT and SD-OCT.
- To assess the clinical significance of differences in segmentation for monitoring disease progression.
- To evaluate the impact of switching OCT devices on patient monitoring.
Main Methods:
- A prospective cross-sectional study involving glaucoma patients and healthy controls.
- Macular scans were acquired using both SD-OCT and HR-OCT on the same eyes.
- Automated segmentation analyzed the thickness of six retinal layers across 97 B-scans.
Main Results:
- Mean differences in retinal layer thickness between SD-OCT and HR-OCT ranged from -2.65 to 2.02 μm.
- Hyper-reflective layers appeared thicker with SD-OCT, while hyporeflective layers appeared thicker with HR-OCT.
- Intra-device differences were within ±1 μm; image quality did not significantly predict inter-device differences.
Conclusions:
- Small but consistent inter-device differences exist between SD-OCT and HR-OCT retinal layer thickness measurements.
- These differences, particularly the alternating pattern, may be due to variations in axial resolution and border definition.
- Switching between SD-OCT and HR-OCT for patient follow-up requires careful consideration of potential masking or masquerading of disease progression due to measurement discrepancies.
Related Concept Videos
Spectroscopy of Carboxylic Acid Derivatives
Infrared spectroscopy is primarily used to determine the types of bonds and functional groups. In carboxylic acid derivatives, a typical carbonyl bond absorption is observed around 1650–1850 cm−1. For esters, the absorption is recorded at around 1740 cm−1, while acid halides show the absorption at about 1800 cm−1. Another acid derivative, the acid anhydrides, exhibit two carbonyl absorption around 1760 cm−1 and 1820 cm−1, arising from the symmetrical and unsymmetrical carbonyl vibration.
In the...
In the...
Attenuated Total Reflectance (ATR) Infrared Spectroscopy: Overview
Attenuated total reflectance (ATR) infrared spectroscopy is a powerful analytical technique used to study the composition of materials. It is widely employed in chemistry, materials science, forensic science, and other fields where sample characterization is required. ATR has several advantages over traditional transmission IR spectroscopy, including the requirement of little to no sample preparation and the ability to analyze a wide range of samples.
The ATR process begins by directing a beam...
The ATR process begins by directing a beam...