Related Experiment Video
Updated: Jul 15, 2026

Endoscopic Ultrasound-Guided Biliary Drainage: Endoscopic Ultrasound-Guided Hepaticogastrostomy in Malignant Biliary Obstruction
Published on: March 25, 2022
Guidewire Obstruction of Centrifugal Pump, in Patient on Extracorporeal Circulatory Support
Johannes Bladt Andersen1, Henrik Frederiksen Højgaard1, Claus Andersen1,2
1Department of Cardiothoracic and Vascular Surgery, Odense University Hospital, Odense, Denmark.
Abstract:
This case report tells the story of a 36-year-old overweight male with suspected myocarditis, who goes into cardiac arrest, and is subsequently transported to a heart center for extracorporeal membrane oxygenation. Mechanical support was established and running until distal limb perfusion was attempted. It is suspected that an aspired guidewire of a central venous catheter placed with the Seldinger technique was aspired and eventually obstructed the impeller of the centrifugal pump.