WebCalEM: a browser-based tool for routine and accurate pixel-size calibration in cryo-EM
Lan Dang1, Zhiqing Wang2, Sung Hyun Joseph Cho2
1Department of Biochemistry and Molecular Biology, Pennsylvania State University, University Park, State College, PA 16802, USA.
Summary
WebCalEM simplifies cryo-electron microscopy calibration. This web application makes accurate pixel size determination accessible for routine quantitative analysis, overcoming previous workflow complexities.
Area of Science:
- Cryo-electron microscopy
- Image analysis
- Scientific instrumentation
Background:
- Accurate pixel size determination is crucial for quantitative cryo-electron microscopy (cryo-EM).
- Existing calibration methods are often under-utilized due to installation barriers and complex workflows.
- Routine adoption of calibration procedures is hindered by practical implementation challenges.
Purpose of the Study:
- To develop an accessible web-based application, WebCalEM, for routine cryo-EM pixel size calibration.
- To transform specialized calibration procedures into an accessible practice for core facilities.
- To lower the barrier for documented per-dataset magnification calibration in cryo-EM.
Main Methods:
- Developed WebCalEM, a self-contained HTML file running client-side in any modern web browser.
- Users upload micrographs of specimens with known crystalline lattices (e.g., gold, graphene oxide).
- Real-time visualization and downloadable statistical outputs are provided without server-side computation.
Main Results:
- WebCalEM processes micrographs entirely client-side, suitable for isolated workstations.
- Cross-standard consistency was achieved between gold and graphene oxide measurements.
- Measurements across two microscopes and ten magnifications showed a Bland-Altman bias of -0.005% with limits of agreement [-0.30%, +0.29%].
Conclusions:
- WebCalEM effectively lowers practical barriers to magnification calibration in routine cryo-EM.
- The application facilitates accessible and documented per-dataset calibration.
- This tool promotes wider adoption of essential quantitative analysis practices in cryo-EM.