Method for correcting the temperature dependence of field-type glass electrode pH sensors
Kei Okamura1,2, Takuroh Noguchi3, Mayumi Hatta3
1Faculty of Agriculture and Marine Science, Kochi University, B200, Monobe, Nankoku, Kochi 783-8502, Japan. okamurak@kochi-u.ac.jp.
Abstract:
Seawater pH measurement is essential for monitoring ocean acidification and its biogeochemical impacts. Glass electrode pH sensors are widely used for in situ applications due to their ease of operation, low power consumption, and no reagent requirements. However, their accuracy can be affected by temperature-dependent effects, particularly those associated with the internal reference solution. Conventional calibration methods based on the Nernst response do not explicitly account for temperature-induced variations in the internal solution, which may lead to systematic bias under conditions with large temperature gradients. In this study, a correction method is proposed to address temperature-dependent errors by modeling the internal pH (pHin). The model incorporates two primary temperature-dependent contributions: the intrinsic temperature dependence of the phosphate buffer and potential shifts caused by activity changes in saturated KCl. These effects are combined through the electrode response factor (fR), defined as the ratio of the observed electrode slope to the theoretical Nernst slope, to estimate pHin as a function of temperature and correct measured pH values to a reference temperature of 25 °C. Application of the method to vertical seawater profiles demonstrated a reduction in deviation from colorimetric reference measurements. The corrected results showed reduced temperature-dependent bias across depth. This approach provides a practical framework for improving the reliability of in situ glass electrode pH measurements under varying thermal conditions. The proposed method addresses equilibrium temperature-dependent effects of the internal solution but does not explicitly account for transient hysteresis, pressure effects, or long-term sensor drift.
More Related Videos
09:17Reductive Electropolymerization of a Vinyl-containing Poly-pyridyl Complex on Glassy Carbon and Fluorine-doped Tin Oxide Electrodes
Published on: January 30, 2015
10:01Thermal Scanning Conductometry (TSC) as a General Method for Studying and Controlling the Phase Behavior of Conductive Physical Gels
Published on: January 23, 2018
Related Concept Videos
Potentiometry: Types of Electrodes
The Standard Hydrogen Electrode (SHE) is a widely used reference electrode that maintains zero potential across all temperatures. However, its need for a continuous hydrogen gas supply renders it impractical for everyday use.
An alternative to SHE is the Saturated Calomel Electrode (SCE). This electrode features an...
Potentiometry: Membrane Electrodes
Electrodes: Overview
There are two main types of electrodes in electrochemical cells. The first type, known as the working or indicator electrode, has a potential that is sensitive to the analyte's concentration and reacts to changes in the...
Concentration Cells
Consider the following voltaic cell: