Iridium-192 high-dose-rate calibration discrepancies: A growing concern?
Paul Simpson1, Emily Debrot1, Neisha Hartfield2,3
1Radiation Oncology, Calvary Mater Newcastle, NSW, Australia.
Journal of Contemporary Brachytherapy
|July 23, 2026
Summary
A growing difference between measured and manufacturer-reported reference air kerma rates (RAKR) for high-dose-rate (HDR) iridium-192 sources was found across multiple countries. This suggests a widespread issue, not a local anomaly, impacting brachytherapy accuracy.
Area of Science:
- Medical Physics
- Radiation Oncology
- Nuclear Medicine
Background:
- Discrepancies in reference air kerma rate (RAKR) for high-dose-rate (HDR) iridium-192 (192Ir) sources have been observed.
- This issue was initially noted at a single Australian hospital, raising questions about its prevalence.
Purpose of the Study:
- To investigate if the observed RAKR discrepancy is a localized event or a more widespread problem.
- To assess the trend of RAKR measurements from 192Ir HDR brachytherapy sources over time.
Main Methods:
- A survey was distributed to brachytherapy physicists in Australia, New Zealand, the UK, France, and Ireland.
- Data collected included HDR 192Ir source details, well chamber models, calibration providers, and historical RAKR measurement discrepancies.
- Linear regression analysis was applied to data from 32 clinics, encompassing 1150 observations over 2 to 20 years.
Main Results:
- A consistent trend of increasing local RAKR measurements compared to manufacturer certificate values was observed across all analyzed series.
- Clinics using well chambers calibrated at NPL or UWADCL showed an average annual discrepancy increase of approximately 0.2% from 2015 onwards.
- All analyzed 192Ir HDR sources were manufactured by Curium.
Conclusions:
- There is an increasing discrepancy between locally measured and manufacturer-certified HDR 192Ir source strengths.
- The manufacturer's measurement procedures may not align with user standards, necessitating a review.
- Brachytherapy societies may need to re-evaluate the requirement for direct comparison between local and manufacturer measurements if agreement does not improve.