Adsorption characteristics of erythropoietin in two β2-microglobulin absorption systems: a preliminary in vitro study
Shoichiro Hara1, Masakazu Hirose2, Taisuke Goto2
1Medical Corporation Imamura Clinic, 2-2-1 Kongo Yahatanishiku, Kitakyushushi, Fukuokaken, 807-1263, Japan. s.hara.imacli@gmail.com.
Purpose:
To compare the erythropoietin adsorption characteristics of two adsorption devices, Lixelle® and FILTOR®, used for the treatment of dialysis-related amyloidosis, using bovine serum.
Methods:
Three recirculating perfusion studies were performed using bovine serum containing recombinant human erythropoietin (epoetin beta; EPO), darbepoetin alfa (DA), or epoetin beta pegol (C.E.R.A.). Because direct quantitative assays for DA and C.E.R.A. were unavailable, and it was difficult to obtain suitable measurement systems for these agents, they were evaluated using an erythropoietin immunoassay based on assay cross-reactivity.
Results:
In the EPO perfusion test (n = 3), the 4-h EPO removal rate was 18.7 ± 3.2% in the Lixelle® group and 53.0 ± 3.1% in the FILTOR® group (p < 0.05). In the DA perfusion test (n = 3), the 4-h removal rate based on erythropoietin immunoassay measurements was 3.3 ± 1.5% in the Lixelle® group and 31.0 ± 6.1% in the FILTOR® group (p < 0.05). In the C.E.R.A. perfusion test (n = 3), the corresponding removal rates were 0.3 ± 0.6% and 1.3 ± 2.3%, respectively, with no significant difference (p = 0.5351).
Conclusions:
The FILTOR® adsorption device showed greater reductions in erythropoietin concentration than the Lixelle® device in experiments using EPO. A similar trend was observed for DA, whereas no apparent difference was observed for C.E.R.A. Because DA and C.E.R.A. were evaluated using assay cross-reactivity, and quantitative assessment was limited by the sampling strategy, these findings should be interpreted with caution.
More Related Videos
09:24Synthesis, Hemoglobin Encapsulation and Biorthogonal PEGylation in Hierarchically Porous UiO-66 Nanoparticles for Oxygen Delivery Applications
Published on: May 8, 2026
08:53A Comprehensive Pipeline to Assess the Efficiency of Human Erythropoiesis In Vitro and Ex Vivo
Published on: January 10, 2025
Related Concept Videos
Methods for Studying Drug Absorption: In vitro
The diffusion cell method uses a two-compartment cell, including a donor compartment with the drug solution, which simulates the environment where the drug is applied, and a receptor compartment with a buffer solution, which simulates the environment...
Methods for Studying Drug Absorption: In situ
The Doluisio method involves perfusing a prepared segment of a rat's small intestine with a solution of radiolabeled drug and a non-absorbable marker. This helps to differentiate between absorbed and non-absorbed drug concentrations. The intestinal segment is connected at both ends using tubing and syringes,...
Extracorporeal Removal of Drugs: Hemoperfusion and Hemofiltration
Drug Elimination by Renal Route: Tubular Secretion
Factors Influencing Drug Absorption: Presystemic Elimination