Integrated Compensatory Responses in a Human Model of Hemorrhage
Published on: November 20, 2016
Real-Time Comparison of Machine Learning-Enabled Devices for Measuring Compensatory Reserve Status
Carlos Bedolla1, Jose M Gonzalez1, Ryan Ortiz1
1Expeditionary Medical Systems Department, U.S. Army Institute of Surgical Research, JBSA Fort Sam Houston, San Antonio, TX 78234, USA.
Bioengineering (Basel, Switzerland)
|July 28, 2026
Summary
The compensatory reserve measurement (CRM) device shows results similar to the FDA-cleared compensatory reserve index (CRI) device. This real-time monitoring of compensatory reserve metrics is valuable for prehospital trauma care.
Area of Science:
- Physiology
- Biomedical Engineering
- Medical Devices
Background:
- Central hypovolemia is a critical condition requiring accurate monitoring.
- Existing methods for assessing compensatory reserve may have limitations in real-time application.
- Noninvasive photoplethysmography offers a potential avenue for continuous physiological monitoring.
Purpose of the Study:
- To compare the real-time performance of a novel compensatory reserve measurement (CRM) device against an established compensatory reserve index (CRI) device.
- To validate the CRM device using data from a simulated central hypovolemia study.
- To assess the potential of the CRM device for early detection of compensatory status changes.
Main Methods:
- A clinical research study involving simulated central hypovolemia induced by lower body negative pressure (LBNP).
- Data acquisition using noninvasive photoplethysmography sensors from 20 participants.
- Comparison of CRM and CRI algorithmic outputs using median error, median absolute error, and pooled Pearson correlations.
- Bland-Altman analysis to assess bias between the two devices.
Main Results:
- Strong linear agreement between CRM and CRI with a pooled correlation R² = 0.859.
- Near-zero mean bias (0.13%) between the CRM and CRI devices, indicating high concordance.
- CRM provided early measurement times (mean 20.52 min) comparable to CRI (mean 17.92 min).
Conclusions:
- The compensatory reserve measurement (CRM) device demonstrates statistically similar performance to the FDA-cleared compensatory reserve index (CRI) device.
- Both devices effectively track compensatory status changes during progressive central hypovolemia.
- Real-time tracking of compensatory changes using CRM holds significant value for continuous monitoring and triage in prehospital trauma settings.