Patient Dimensions From a Single CT Image Provide a Clinically Practical and Accurate Method for Dose Estimation in
Claire Whittaker1, Monique Mayer1, Matthew Hutcheson1
1Small Animal Clinical Sciences, Western College of Veterinary Medicine, University of Saskatchewan, Saskatoon, Saskatchewan, Canada.
Abstract:
Size-specific dose estimate (SSDE) uses linear dimensions measured on a computed tomographic (CT) image to scale the scanner-reported volumetric computed tomography dose index (CTDIvol) and thereby estimate dose. The objectives of this retrospective observational study were to evaluate whether the American Association of Physicists in Medicine (AAPM) geometric size metrics for abdominal CT, when implemented using canine-specific size measurements, correlate with water-equivalent diameter (Dw) in canine patients and to compare institutional SSDEs with human benchmarks. Abdominal CT scans from 30 dogs across three body weight categories (<10, 10-30, and >30 kg) were analyzed. The relationships between effective diameters calculated using four geometric metrics-dorsoventral dimension (DV), lateral dimension (LAT), the sum of DV and LAT, and the square root of their product-and Dw, which incorporates both size and attenuation, were analyzed using linear regression. The strongest correlation with Dw was observed for the combined DV and LAT metrics (R2 0.94, p < 0.001). The displayed CTDIvol underestimated SSDE by 50%. The 75th percentile of the SSDE for a single-phase, pre-contrast abdominal examination was 21 mGy, two to four times higher than the US human dose reference level. In conclusion, geometric size metrics measured from a single CT image were in close agreement with attenuation-based SSDE (Dw) for canine abdominal CT scans, supporting the feasibility of applying the AAPM SSDE estimation method to canine abdominal CT imaging. These findings demonstrate the importance of incorporating patient size into CT dose estimation and suggest that institutional SSDE values may warrant dose optimization in veterinary abdominal CT practice.
More Related Videos
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
09:21Human Brown Adipose Tissue Depots Automatically Segmented by Positron Emission Tomography/Computed Tomography and Registered Magnetic Resonance Images
Published on: February 18, 2015
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography