Related Experiment Video
Updated: Aug 5, 2026

Three-Dimensional Phase Resolved Functional Lung Magnetic Resonance Imaging
Published on: June 21, 2024
BART Online Open-Source Sequence Toolbox for Computational MRI
Daniel Mackner1, Philip Schaten1, Markus Huemer1
1Institute of Biomedical Imaging, Graz University of Technology, Graz, Austria.
Purpose:
In advanced computational MRI techniques, acquisition and reconstruction techniques are jointly designed. For reproducibility, it is therefore important to provide an open implementation of both. At the same time, any use in a clinical environment usually requires a close integration with the MRI scanner. Ensuring long-time reproducibility and maintenance then poses additional challenges. In this work, we aim to provide a fully integrated open-source framework that can meet these demands.
Methods:
A software framework to develop pulse sequences is added to the BART toolbox. In addition, a vendor-specific driver sequence is developed that can be used to run the sequence on a clinical MRI scanner enabling online adjustment of all relevant sequence parameters. Using the Pulseq format, the exact same sequence can also be reproduced offline. As proof-of-concept, quantitative MRI methods for and joint water/fat mapping using radial FLASH and model-based reconstruction are implemented in the proposed framework. Consistency between online and offline acquisition is validated in phantom and in vivo experiments.
Results:
Quantitative MRI methods consisting of acquisition and reconstruction were successfully implemented in BART. Acquisition parameters and FOV can be adapted online on a clinical MRI system. Quantitative parameter maps from model-based reconstruction agree for online and offline regenerated Pulseq acquisitions.
Conclusion:
This work enables reproducibility of advanced computational MRI methods within a comprehensive end-to-end open-source framework.