Related Experiment Video
Updated: Aug 12, 2026

Magnetic Resonance Imaging of Multiple Sclerosis at 7.0 Tesla
Published on: February 19, 2021
Head bandaging for patients with auditory implants undergoing MRI, highlighting the risk of magnetic puck projectile
Sean Daniel1, Sonia Kumar1, Rajeev Mathew1
1ENT Department, John Radcliffe Hospital, Oxford, UK.
Abstract:
We present a case of unintended displacement of a manufacturer-provided external magnetic puck (splint), which occurred during an MRI scan in a patient with an auditory implant, highlighting the associated risks and potential for harm. Despite established safety protocols and secure head bandaging, the strong magnetic field resulted in the puck becoming a dangerous projectile, posing a potentially serious risk to both the patient and MRI staff. This report discusses the rationale for the new magnetic puck introduced by the auditory implant manufacturer to try to reduce internal implant magnet displacement. We offer an alternative technique to minimise the risk of splint slippage during head bandaging with non-magnetic standard plastic splints, with the aim of mitigating the risk of internal magnet displacement whilst avoiding the use of this magnetic puck with the potential risks of a magnetic projectile in the MRI scanner.
More Related Videos
Related Concept Videos
Imaging Studies for Cardiovascular System IV: CMRI
Magnetic Resonance Imaging
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies IV: Magnetic Resonance Imaging